Troubleshooting
seca
03.04.2013/JEG
15 (24)
30-34-00-864
3 Troubleshooting
The following section contains an overview of possible error symptoms, their cause and
the steps required to remedy them. The error messages generated by the device itself
are also described below as well as the options for rectifying them.
3.1 Error symptoms scale
Description of
error
Possible causes
Remedy
Power supply defective
Check batteries and/or
power pack (if present)
Voltage supply defective
Check cabling (see
section 4)
Scale does not start
Keypad defective or not
connected
Check cable
connections, check
cabling and keypad for
disconnection, replace if
necessary
Packaging / transport lock not
completely removed
Check scale
Scale not calibrated
Recalibrate scale (see
section 2)
Load cell damaged
Replace load cell
Scale either
showing wrong
weight or none at all
Application of force to load
cell interrupted
Check load cell
connections
Replace electronics unit
Wide variation in
measured values
Electr. measuring unit/load
cell damaged
Replace load cell