this may reduce the life of the unit.
15.4 Touch Current
Test Voltage
Mains supply 230V +10% / -15%
Max Load Current
13A
Input Impedance
2K
Ω
As per IEC 61010-1 A.1
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail Limit
User Defined
Display Range
0.00 – 3.50mA
Measuring Range
0.10 – 3.50mA
Resolution
0.01mA
Accuracy
± 10% ±2 digits
Attention: Avoid prolonged, repeated use at
full load (13A) and excessive test duration as
this may reduce the life of the unit.
15.5 Voltage/Frequency Measurement
Incoming mains voltage and frequency are displayed
during the 2 powered tests only.
Voltage Display Range
0 – 300V
Voltage Measuring Range
20 – 300V
Voltage Resolution
1V
Voltage Accuracy
± 10%
Frequency Measuring Range 50Hz or 60Hz
Frequency Accuracy
Indication only
15.6 Earth Loop Resistance
Test Voltage
Mains supply 230V +10% / -15%
Test Current
Non-trip for 30mA RCD
Test Duration
Approx 10 seconds
Pass/Fail Limit
User Defined
Display Ranges
0.0 – 199
Ω
Measuring Range
1.0 – 199
Ω
Resolution
0.1
Ω
, 0.0 – 19.9
Ω
1
Ω
, 20 – 199
Ω
Accuracy
± 10% ±2 digits
15.7 RCD Trip Time
Test Voltage
Mains supply 230V +10% / -15%