PrimeTest 300
Operating Instructions
- 51 -
Press
Class
(F2) to set the class of the appliance
under test. For a class II appliance the probe symbol
will be displayed at the top of the LCD display.
For both Class 1and Class II appliances fit the
Appliance mains plug into the EUT outlet. For Class 2
appliances only connect the Earth Bond lead to the
appliance. Press the green
START
key to begin the
test. The test will run for the duration set or until the red
STOP
key is pressed.
11.3 Protective Earth (PE) Conductor Current
Refer to Fig 5
This test is applicable on Class I
Equipment.
The Protective Earth (PE) Conductor Current Test
displays the current flowing in the earth conductor of
the EUT. This is measured by looking at the difference
in current flowing in the live and neutral conductors.
The Tester displays the result in milliamps (mA).
In the top left area of the screen, the tester also
displays the EUT load in kVA.
Warning: Always test the earth continuity
and insulation resistance before performing
a PE Conductor Current test.
Warning: Always check that an appliance
with moving parts (e.g. an electric drill) is
safely mounted to avoid risk of damage to
equipment or personnel.