Interfacing
44
Interfacing
The Analyser provides a single 9 way connector for a serial interface to Rigel Medical’s optional serial printer.
Note: Rigel Medical only recommend the use of Rigel Medical approved printers. The use of alternative
printers to those supplied by Rigel Medical may impair the performance of the Rigel 266 plus.
Connection to other earthed periphirals (ie laptops and PC’s) will effect the reading of the Rigel 266
plus.
An option Blue Tooth RS 232 adaptor is allow for real time download to earthed peripherals (ie Laptops & PC’s).
Please contact our sales department for further information.
When using the Rigel Medical Thermal Printer (model MCP9810) the printer’s DIP switch settings should be set
to SW1-3 on, SW4 off, SW5-8 on. Ensure the printer is fully charged before use with the
Rigel 266 plus
as
erroneous operation may result. See separate MCP9810 printer manual for details.
The Connector pin-outs for the Serial port are shown below for information only:-