Patient Leakage - Applied Part To Earth

 

 

28

EUT Power Break 

5. rEV 

4. rEV 

Reversed 

None 

6. rE 

Reversed 

Earth Open 

 

Note: Tests marked * are not performed for Class II equipment. 

 

If the unlimited test duration is selected the user must press the Start key to stop the test.  To initiate the next 
test in the sequence the Start key must be pressed again.  In this mode the Analyser can be used to test 
equipment with slow start-up speeds. 

 

 

 

 

 

 

 

 

 

 

 

 

For type B and BF equipment, the patient leakage current is measured with all parts of the applied part 
connected together, shown dotted above. For type CF equipment the patient leakage current is measured from 
each part of the applied lead separately. IEC601-1 amendment 2 specifies separate maximum allowed values 
for AC and DC current - both of which are measured by the Safety Analyser. 

Patient Leakage, normal condition 

 

This test measures the patient leakage current under normal conditions. The current is measured through the 
Measuring Device with S1 and S8 closed, S5 normal and then S5 reversed. The maximum allowed values are 
as follows: 

 

 

Type B 

Type BF 

Type CF 

Class I   DC 

0.01 mA 

0.01 mA 

0.01 mA 

              AC 

0.1 mA 

0.1 mA 

0.01 mA 

Class II  DC 

0.01 mA 

0.01 mA 

0.01 mA 

              AC 

0.1 mA 

0.1 mA 

0.01 mA 

 

 

Patient Leakage, single fault, supply open 

 

This test measures the enclosure leakage current with a single fault condition (supply open).  

Summary of Contents for Rigel 266 Plus

Page 1: ...Rigel 266 Plus Compact Electrical Medical Safety Analyser Instruction Manual 350A555 Issue 1 0 February 2006 2006 Seaward Electronic Ltd Issue 1 0...

Page 2: ......

Page 3: ...ically mechanically by photocopying recording or in any manner without prior written consent from SEAWARD Electronic Limited This also applies to accompanying drawings and diagrams Due to a policy of...

Page 4: ...nt Type 17 Earth Continuity Tests 17 Insulation Tests 18 Powered Tests 20 Load Test 20 Earth Leakage Test 22 Earth Leakage normal condition 23 Earth Leakage single fault supply open 23 Enclosure Leaka...

Page 5: ...YSER 46 Cleaning the Analyser 46 User Maintenance 46 CHAPTER 6 ACCESSORIES 48 Standard Accessories 48 Optional Accessories 48 CHAPTER 7 SPECIFICATIONS 49 AC Earth Bond Test 49 Insulation Test 49 Earth...

Page 6: ...onforms with Electromagnetic Compatibility Directive 89 336 EEC and Low Voltage Directive 73 23 EEC Standards used EN 60601 1 1990 Medical Electrical Equipment General Requirements for Safety EN 61010...

Page 7: ...icate 2 Read the operating instructions fully before conducting any tests 3 Contact Rigel Medical if you require training on this Analyser or any other Rigel Medical product Courses can be arranged at...

Page 8: ...he user is familiar with the instructions covering the capabilities and operation of this instrument The user should ensure that all reasonable safety precautions are followed and if any doubt exists...

Page 9: ...is manual is read fully before proceeding with any tests Should the Analyser behave abnormally do not continue with the testing Disconnect immediately and contact Rigel Medical for servicing see Chapt...

Page 10: ...e provided with innovative features to aid difficult test situations which allow testing of a wide variety of equipment The Analyser combines state of the art testability with the following innovative...

Page 11: ...our compact medical safety analyser Item Number Part 1 Rigel 266 plus EMSA 2 Carrying case 3 Earth Bond Lead 4 Power cord Integral with the Rigel 266 plus 5 Applied Part Lead Set including Applied Par...

Page 12: ...Keyboard Item Part 1 Rotary Test Selection Switch 2 Semiautomatic Mode Up Down Cursor keys Manual Mode UP SFC Neutral Open DOWN SFC Earth Open DOWN SFC Mains on AP 3 Semiautomatic Mode SELECT button...

Page 13: ...he tester performs the SFC test automatically Manual Mode The User initiates the SFC tests Applied Part A part of the EUT that in normal use comes into contact with a patient SFC Single Fault Conditio...

Page 14: ...rranged on the switch such that a clockwise rotation of the switch needed to perform a normal sequence of tests Arrow keys and allow for adjustment of test variables e g of the test time The Select ke...

Page 15: ...on of the Analyser status a primary 4 digit 7 segment display along with a 4 digit secondary display and a 2 digit duration time display Test Icons Indicates the test selected Insulation Patient Auxil...

Page 16: ...est sequence After a test this display compares the test result with standard limits During the leakage tests this displays the test being carried out in the test sequence This is indicated as a text...

Page 17: ...Connecting the Analyser 15...

Page 18: ...rent tests Earth Leakage Enclosure Leakage Patient Leakage Patient Leakage F Type Mains on AP Patient Leakage SIP SOP and Patient Auxiliary Current Rotate the switch to the appropriate leakage test Se...

Page 19: ...ng the SELECT key will cause the equipment type to stop flashing indicating that the equipment class and type had been selected During tests where the equipment class and type is displayed the operato...

Page 20: ...he mains supply plug and the earth bond test lead clip probe A high current is normally used to stress the connection under fault conditions The length of the test should be limited to prevent damage...

Page 21: ...selected The Analyser displays the insulation resistance measured The Analyser can test either the EUT by itself or all the Applied Parts internally connected together Press the SELECT button and the...

Page 22: ...sage appears allowing the user to continue This message will be displayed if a Live to Neutral short exists or the EUT mains current consumption is too high and if tests are continued the Analyser s f...

Page 23: ...ets with the rated voltage of that socket The Analyser measures the power used by the appliance and displays the reading in kVA Power to the appliance can be set to remain on until the rotary switch i...

Page 24: ...conditions neutral or earth open circuit is done automatically during the test The current test sequence step is shown on the lower display Should a test fail the Analyser will stop the automatic tes...

Page 25: ...ion This test measures the earth leakage current under normal conditions The current is measured through the Measuring Device with S1 closed and S5 normal and then S5 reversed The maximum allowed valu...

Page 26: ...uched the appliance This is based on the standard IEC601 body model The Analyser uses the earth bond probe to detect any current flowing on the appliance metalwork attached to an appropriate point on...

Page 27: ...bond probe is used to make contact with all conductive non protectively earthed parts of the equipment Enclosure Leakage normal condition This test measures the enclosure leakage current under normal...

Page 28: ...pen This test measures the enclosure leakage current with a single fault condition earth open The current is measured through the Measuring Device with S1 closed S8 open and S5 in normal and then S5 r...

Page 29: ...detects any current flowing in the Applied Parts Lead The Analyser displays the result in milliamps mA To change the test duration press the SELECT key and then UP or DOWN keys The user can also sele...

Page 30: ...type CF equipment the patient leakage current is measured from each part of the applied lead separately IEC601 1 amendment 2 specifies separate maximum allowed values for AC and DC current both of whi...

Page 31: ...ss II n a n a n a Patient Leakage F Type Switch Position PATIENT F TYPE Test Icon Displayed Warning This test involves applying a current limited mains potential 110 of mains input voltage to the Appl...

Page 32: ...user can toggle between different Patient Connections as required to take the individual measurements The selection of normal or reverse operation and single fault conditions neutral or earth open ci...

Page 33: ...en conducting this test Current limiting is via a limiting resistor in series with the SIP SOP socket NOTE The Earth bond probe should not be used as part of these tests as this may affect the current...

Page 34: ...nt the Single Fault Earth Open tests are skipped The test sequence is shown below Test Sequence Mains Single Fault 1 nS Normal Mains on SIP SOP Normal 2 nSr Normal Mains on SIP SOP Reversed 3 rS Rever...

Page 35: ...hown dotted above The current is measured through the Measuring Device with S1 closed S8 closed and S5 and S9 normal and then S9 reversed The maximum allowed values are as follows Type B Type BF Type...

Page 36: ...ient Connection or Applied Part prior to commencing the test When the test duration is set to UL unlimited the user can toggle between different Patient Connections as required to take the individual...

Page 37: ...erated in unlimited test duration mode In this mode of operation the Start key must be pressed to stop the current test and pressed again to initiate the next test in the test sequence In this manner...

Page 38: ...Device with S1 and S8 closed S5 normal and then S5 reversed The maximum allowed values are as follows Type B Type BF Type CF Class I DC 0 01 mA 0 01 mA 0 01 mA AC 0 1 mA 0 1 mA 0 01 mA Class II DC 0 0...

Page 39: ...mA 0 05 mA 0 05 mA AC 0 5 mA 0 5 mA 0 05 mA Class II n a n a n a IEC Lead Polarity Test Switch Position POLARITY Test Icon Displayed Warning 40Va c test voltage applied to lead Plug the lead into the...

Page 40: ...C mains polarity and AC DC parameter before initiating a test The user presses the START to begin the test Power is maintained to the EUT even when the user changes the SFC or when an Applied Part is...

Page 41: ...on SIP SOP AP Earth Yes No No No No Enclosure Yes Yes No No No Patient Yes Yes No No Yes Patient F Type No No Yes No Yes Patient SIP SOP No No No Yes Yes Patient Auxiliary Yes Yes No No Yes The polar...

Page 42: ...Operation 40...

Page 43: ...rection of the fault If a voltage greater than 30V is detected on the neutral input line or the earth is missing then the following display will be seen L N REVERSED Err2 Check for incorrect wiring at...

Page 44: ...Medical for repair Safety Tests During Operation The Analyser performs self tests during normal operation An internal earth bond test and a low voltage test are performed prior to applying mains power...

Page 45: ...nents are not overheated High rates of testing may cause this situation especially with a series of long duration Earth Continuity tests Setting Earth Continuity test duration to 2 seconds will increa...

Page 46: ...PC s will effect the reading of the Rigel 266 plus An option Blue Tooth RS 232 adaptor is allow for real time download to earthed peripherals ie Laptops PC s Please contact our sales department for f...

Page 47: ...facing 45 Serial Port The serial port uses a standard 9 way D type connector Pin Description 1 N C 2 RX 3 TX 4 DTR 5 0V 6 N C 7 0V 8 N C 9 5V Baud Rate 9600 Start Bits 1 Data Bits 8 Stop Bits 2 Parity...

Page 48: ...ll reduce both the life of the instrument and its reliability If the Analyser is subject to condensation allow the Analyser to completely dry before use Always check the Analyser and all test leads fo...

Page 49: ...e department to obtain a Returns Number By obtaining a Returns Number your service request can be booked in advance thus reducing the down time of your equipment When asking for a Returns Number pleas...

Page 50: ...nalyser The standard accessories are supplied with the Analyser Standard Accessories Accessory Part Number Earth Bond Test Lead 286A925 Mains Lead 270A025 Applied Part Adaptor Lead 286A950 Carry Bag 2...

Page 51: ...gits Misc 4 wire measurement floating earth Insulation Test Test Voltage 500V d c nominal 0 5M load Short Circuit Current 2mA d c maximum Range 100k 100 M Resolution 0 01 M 100 M Accuracy 5 of reading...

Page 52: ...ng Resistance Nominal 48K 230V Units Nominal 24K 110 120V Units SIP SOP Open Circuit Voltage 110 20 of Mains Input Voltage Patient F Type Test Input Impedance 1Mohm Frequency Response as per IEC 60601...

Page 53: ...C non condensing Storage 10 C to 50 C non condensing Maximum R H 90 Supply Rating UK European Australian Versions 230V 10 50 60Hz US Version 115V 10 50 60Hz Power Rating 30VA Max EUT Output Current UK...

Page 54: ...mA 1mA 0 5mA 1mA Enclosure Leakage 0 1mA 0 5mA 0 1mA 0 5mA 0 1mA 0 5mA Patient Leakage dc 0 01mA 0 05mA 0 01mA 0 05mA 0 01mA 0 05mA Patient Leakage ac 0 1mA 0 5mA 0 1mA 0 5mA 0 01mA 0 05mA Patient Lea...

Page 55: ...BS3456 Household electrical appliances 0 1 R BS4533 Luminaires 0 5 BS2769 Hand held motor operated tools 0 1 R BS415 Mains operated electronic and related apparatus 0 5 R BSEN 60950 Information techno...

Page 56: ...15 Mains operated electronic and related apparatus 0 5 1 0 BSEN 60950 Information technology equipment 0 5 1 0 Information technology equipment not constructed to BSEN 60950 May be damaged by insulati...

Page 57: ...844 2 may be obtained from The Institution of Electrical Engineers P O Box 96 Stevenage Herts SG1 2SD England Copies of the Medical Devices Agency Checks and tests for newly delivered medical devices...

Reviews: