5
.Cleaning the inside of drawer chamber
Step 1: Clean the inner wall of drawer chamber
Use the cleanroom swabs dampened with 75% isopropyl alcohol, and then put swabs into
the chamber. Clean the inner war from right to left with swabs, repeated gently until the
inner wall has no obvious dirt or the head of swabs is clean.
CAUTION: When cleaning, do not touch the temperature sensor at the right side showing
on picture.
It is crucial to clean the location 3 well showing in the picture.
Step 2: Clean the key position 1
Use the cleanroom swabs dampened with 75% isopropyl alcohol, and then put swabs to the
cleaning position 1, according to the direction of arrow showed in picture to gently wipe back
and forth until the inner wall has no obvious dirt or the head of swabs is clean.
Step 3:
Clean the key position 2 and 3
Use the cleanroom swabs dampened with 75% isopropyl alcohol, and then put swabs into
the chamber to clean position 2 and 3, according to the direction of arrow showed in picture
to gently wipe back and forth until the inner wall has no obvious dirt or the head of swabs is
clean.
Instructions
do not touch the temperature
sensor at the right side showing
on picture