Please read this manual carefully before using analyzer.

SMT-120V

User

 

Manual

 

 Fully Automated Veterinary 

Chemistry Analyzer

Summary of Contents for SMT-120V

Page 1: ...成 都 斯 马 特 科 技 有 限 公 司 Please read this manual carefully before using analyzer SMT 120V UserManual Fully Automated Veterinary Chemistry Analyzer ...

Page 2: ...4 1 Frequently used buttons 4 1 1 Setting 4 1 2 Sample data 4 1 3 QC data 4 1 4 Preparing analysis 4 2 Soft keyboard Section 2 Installation of Analyzer Section 3 Turning the Analyzer On and Off Section 4 Operation 19 19 20 20 21 22 25 26 26 26 28 28 28 29 30 5 1 Sample requirements 5 2 Preparing the reagent disc 5 2 1 Storage and preparing 5 2 2 Dispense the sample 5 3 Sample test 5 4 Test process...

Page 3: ... 1 3 Cleaning the inside of drawer 6 1 4 Comparison of before and after cleaning 6 2 Updating the analyzer software 6 2 2 Updating by WIFI 6 3 Troubleshooting 6 4 Consumables replacement Section 7 Packing Storage andTransportation Section 8 Contact Information 5 6 2 2 QC test report 5 6 2 1 Sample test report Section 6 Maintenance Service 6 2 1 Updating by USB flash drive ...

Page 4: ...ble information in this manual but it is inevitable to have mistakes or omission If you have any doubts please kindly contact us to specify them Contact information has been provided in the last page of this manual In order to improve the performance and reliability of the analyzer we will make some real time changes to the hardware or software which may lead to the difference from this manual Sea...

Page 5: ...nalyzer is exactly the same as the packing list 1 If you have any questions please contact our salesperson or distributor Please be sure to fill in the guarantee card and send product warranty receipt back to 2 us as soon as possible If you email us please send it to sales seamaty com Please read all documents carefully and save them for future need 3 Warning The analyzer should be protected from ...

Page 6: ...e grounding well Do not place this analyzer in a position where it is difficult to disconnect the switch The production date of the analyzer is shown on the label at the back of the analyzer The reagent discs in this manual are Pulitai reagent discs which match with SMT 120V chemistry analyzer In accordance with the standard laboratory practice consider the potential biological risk of animal bloo...

Page 7: ...oves during operation or else there is a potential risk of biological infection There is also a potential biosafety risk in the absence of biosafety treated of used reagent disc Once the skin is in direct contact immediately wash and disinfect the contact area and consult a doctor Operating reagent disc 4 Reagent beads may contain corrosive substances please follow the instructions in accordance w...

Page 8: ...or only need to collect blood samples heparin lithium anticoagulation of whole blood plasma or serum with 90 120 μl added into the reagent disc rotor and then insert the rotor into analyzer to accomplish the testing automatically with immediately printing out result In order to prepare for a second review it is recommended that the operator collects not less than 250 μl of sample at a time There a...

Page 9: ...test and the operating principle is It will produce immune complexes after the antigen antibody binding occurred with turbidity at a certain time When the light passes through the solution it will be absorbed by immune complexes The more the amount of immune complexes the more light is absorbed The intensity of light being absorbed is proportional to the content of immune complexes within a certai...

Page 10: ...lters 340 405 450 505 600 630 850 nm Test method end point rate fixed time etc Heparin lithium anticoagulation whole blood serum and plasma samples The large capacity memory to meet customer needs Intelligent real time quality control function to ensure accuracy External mouse and printer are available based on USB 15V vehicle power supported it is strongly recommended to connect the analyzer afte...

Page 11: ...Sample Type Sample Volume Bar Code Testing Time Testing Principle Testing Method Temperature Absorbance Resolution Carryover QC Calibrate Work Environment Light Source Storage Printer IT Connection Weight Optic System Power Adapter Display Instructions ...

Page 12: ...uirements for power supply AC100 240V 50 60Hz 105W Max The analyzer can not use the same power supply socket with some high power equipment including centrifugal refrigerator and oven etc The power supply should be connected with three cores wire and and good grounding the voltage between the neutral line and earth line 5V or choose two core power cord with power adapter provided by Seamaty 2 2 In...

Page 13: ...e printer 4 P N E00SMT120VCO Version A 0 12 The specification of thermal print paper is 50 57cm the inner structure seeing Figure 2 1 2 2 2 Set up printing paper Figure 2 1 Steps 1 Open the printer cover Remove the packing of printing paper load the paper roll into the printer chamber 2 with correct direction see Figure 2 2 Figure 2 2 3 Pull out the roll paper to cross the printer slot Close and l...

Page 14: ...bright with the screen showed as Figure 3 1 Figure 3 1 Loading and initializing the screen shows as Figure 3 2 Figure 3 2 Turning off Press down the switch button in the back of the analyzer when the drawer is closed disconnect the plug from analyzer the indicator light off Note To extend the lifespan of the analyzer and accessories it required to follow the procedures of powering off and do not f...

Page 15: ...When the loading is completed the home screen will show interface as figure 4 1 Please check the date and time in the status bar and you could reset them according to section 4 1 1 Figure 4 2 This button is used to set system date time hospital name printer mode sample or quality control range etc Click Setting icon in the home screen and enter into setting menu showed as Figure 4 2 4 1 1 Setting ...

Page 16: ...ct analyte in the list to modify the parameters and press Save The analyzer includes a number of factory set analyte and reference parameters but it is available for user to modify thembased on the needs 1 Press System Time icon in the setting screen In the Date Time Setting screen set date time time zone etc Click Done to save 2 your time setting Find Drop Down Menu in the right corner click the ...

Page 17: ... press Save Figure 4 6 It is used to search and review the test results to modify the sample information and to print out test results seeing Figure 4 6 4 1 2 Sample Data Press Query icon to show screen as Figure 4 7 User can search history results by Sample No Name or Date Click Sample Detail icon to view or modify the sample information Click Print icon to print out the result Instructions ...

Page 18: ...modify the sample information and to print out result seeing Figure 4 8 4 1 3 QC Data Press Query icon to show screen as Figure 4 9 User can search history results by Quality Control ID Lot or Date Click QC Detail icon to view or modify the sample information Click Print icon to print out the result Figure 4 9 Instructions ...

Page 19: ... Analysis icon to start testing If there are messages showed on screen read it carefully before operation or contact our after sales service department if any problem occurred 4 1 4 Prepare Analysis The analyzer has a built in soft keyboard Click the space where you want to input text to activate the soft keyboard Figure 4 11 Instructions ...

Page 20: ...in lithium heparin collection tubes turn upside down 3 5 times slightly to ensure the sample and the anticoagulant mixed well Whole blood must be analyzed within 30 minutes of collection or separated into plasma or serum To prevent hemolysis do not refrigerate or shake whole blood If not analyzed immediately plasma or serum can be stored at 2 8 C for no longer than 24 hours after centrifugation or...

Page 21: ... time dispose it instead of putting back to the refrigerator and use it again A disc can remain in its sealed pouch at room temperature about 25 C for a cumulative period of 48 hours Longer time at room temperature can cause suppression of chemistries and disc cancellations Do not expose discs in or out of their foil pouches to direct sunlight or to temperatures above 32 C Inspect the unopened foi...

Page 22: ...d on the disc seeing Figure 5 3 The appropriately value is from 90 to 120 μl More than 120 μl of sample will overfill the chamber Less than 90 μl also will affect the test results Figure 5 3 3 To avoid cross contamination the pipette tip can not be used repeatedly to draw multiple samples Hold the disc by its edges in a flat position and insert it into analyzer 4 Caution 1 Do not contaminate or da...

Page 23: ...of the tissue in a biohazard container Discard the pipette tip into a biohazard container 6 5 3 Sample Testing 1 Booting and preparing 2 Self test and warming up Note The screen will show as Figure 5 4 after booting It need about 5 minutes to warm the disc chamber to operating temperature Figure 5 4 3 Click Prepare Analysis icon the drawer will be open Insert disc into drawer showing as Figure 5 5...

Page 24: ...shows as Figure 5 6 Figure 5 6 4 Press Start Test icon the drawer will be closed and the analyzer will read the two dimensional bar code and enter into testing screen seeing Figure 5 7 Figure 5 7 Note If the analyzer failed to read two dimensional bar code the screen will show a message as Figure 5 8 Figure 5 8 Instructions ...

Page 25: ...sc into drawer again to start test If the bar code has been damaged please do not use it In the testing process user can input information of operator and sample Please follow 6 the steps Press Operator icon to enter into screen as Figure 5 9 Figure 5 9 7 Click the input field when the cursor appeared input the information of Operator Department then click OK seeing Figure 5 10 Figure 5 10 8 Press...

Page 26: ... drawer closed when not using Disassembling the analyzer without permission will void the warranty P N E00SMT120VCO Version A 0 25 Figure 5 11 9 Do not press Stop icon during the testing unless it is in emergency Click Stop it will show a dialog box as Figure 5 12 Figure 5 12 Click OK to stop the testing Note The reagent disc can not be used again as stopped in testing 5 4 Test Process Summary Ins...

Page 27: ...rmation to perform its calibration along with the Real time Qality Control RQC system in analyzer when ever a reagent disc is run In this way the analyte concentrations produced by the analyzer will be accurate 5 5 2 Quality Control The performance of the analyzer can be inspected by running an external quality control The standard substance for quality control is the same as the test sample It is...

Page 28: ...it will show screen as Figure 5 16 User can check or print out the result P N E00SMT120VCO Version A 0 27 Figure 5 13 Start testing with showing the screen as Figure 5 14 Figure 5 14 Edit Operator and QC Detail information Please refer to Section 4 3 5 1 for Operator Click QC Detail to enter into screen as 5 15 Figure 5 15 Instructions ...

Page 29: ...d in four columns 2 assay name analyte concentration reference range and specified units as shown at right The deviation between the test value and the reference range will be followed by H or L standing for higher than or lower than reference range The sample indices are included at the bottom of the results 3 printout WAT means empty cuvettes fills with diluent only as a control on the system cu...

Page 30: ...des information such as hospital name test time QC 1 name QC batch number operator ID number lab and disc lot number The test results section of the card is printed in four columns assay name analyte 2 concentration reference range and specified units The explanation of QC sample indices included at the bottom refers to Section 5 6 5 3 Figure 5 18 Instructions ...

Page 31: ...nutrition tumor and other diseases can cause abnormal index PHOS The increased phosphorus value is the kindney index especially for the late kidney disease Mg Evaluate adrenal gland and kidney function K Provide cell the main buffer environment and help maintain acid base balance and osmotic pressure Na Maintain osmotic pressure of the body and acid base balance and transmission nerve impulse Cl It...

Page 32: ...ure 5 20 Figure 5 20 Instructions HospitalName CONTROL REPORT TEST TIME 2017 08 17 06 56 CONTROL QC LOW QC BATCH NO 9 OPERATOR ID LAB REAGENT BATCH NO 917221 Ver 1 00 00 40 1 00 01 12 Assay Conc Avg Ref Unit tCO2 50 0 mmol L Ca 0 46 mmol L PHOS 0 10 mmol L Mg 0 20 mmol L K 8 00 mmol L Na 80 0 mmol L Cl 50 0 mmol L WAT 0 EMP 0 CHE 0 ...

Page 33: ...er towel with the detergent then apply to the analyzer 6 1 2 Cleaning the display Clean the analyzer s screen periodically using a soft lint free cloth dampened with a glass cleaning fluid or window cleaner 6 1 3 Cleaning the inside of drawer Note Seamaty recommends only the cleaning methods described in this section If for some reason another method is needed contact Seamaty for Technical Support...

Page 34: ...icon the drawer will be opened 2 3 Turning off analyzer CAUTION Only to do maintenance after powering off 4 Insert the clean white paper Insert the clean white paper into the inside of drawer along the surface of the drawer plate until the paper can not move forward the picture showing as below Instructions ...

Page 35: ...he picture Step 2 Clean the key position 1 Use the cleanroom swabs dampened with 75 isopropyl alcohol and then put swabs to the cleaning position 1 according to the direction of arrow showed in picture to gently wipe back and forth until the inner wall has no obvious dirt or the head of swabs is clean Step 3 Clean the key position 2 and 3 Use the cleanroom swabs dampened with 75 isopropyl alcohol ...

Page 36: ...commended period is weekly for a basic cleaning and monthly for a deeper cleaning It needs an immediately cleaning when there is foreign matter or dirt in the chamber caused by accident Picture of before cleaning Picture of after cleaning 6 2 Updating the Analyzer Software Update the analyzer software in the below occasions A When there is an update notice of the new released software form Seamaty...

Page 37: ...rmation Creat a SMT120 folder in the root directory of USB flash drive copy zip 3 and md5 format files into this folder showing as Figure 6 1 CAUTION Updating has risks Please read the update steps or tips carefully and confirm the right software update package and version before upgrading avoidingthe update error which may cause the damage of analyzer Figure 6 1 CAUTION Make sure the upgrade pack...

Page 38: ...s please restart the analyzer If fails again please contact our after sales service team the detail contact information seeing Section 8 6 2 2 Updating by WIFI 1 Prepare a USB wireless adapter seeing Figure 6 2 Note It is recommended to use B link brand with model number BL LW05 AR5 other USB WIFI adapter brands with RT3070 RTL8192CU or RTL8188CUS chip also work well with the analyzer Figure 6 2 1...

Page 39: ...FI seeing Figure 6 5 If there is no WIFI showing in the list unplug the USB WIFI adapter power off and then restart the analyzer reconnect the USB WIFI adapter to analyzer Figure 6 5 4 Connecting network Select a WIFI to connect input account name and password seeing Figure 6 6 Figure 6 6 Instructions ...

Page 40: ...he account name and password matched or not 5 Download the latest software Click Get Versions icon to obtain the software version on the server seeing Figure 6 8 Figure 6 8 Note Be sure the WIFI is connected to internet and the signal is strong 6 Click Download to start the downloading of new version seeing Figure 6 9 Figure 6 9 Instructions ...

Page 41: ... meets the requirement of analyzer 110V 240V AC Restart the analyzer Check if the reagent disc is expired Check if the parameter settings of analytes are right Use USB mouse as an alternative Reconnecting the USB mouse to computer Check if the USB mouse damaged Restart the analyzer Check if the Build in Printer mode has been selected in Setting screen Check if the thermal printing paper is loaded ...

Page 42: ... in build in printer is 50 57mm Reagent Disc Specified Consumable It is recommended only to use reagent discs manufactured by Chengdu Pulitai Biological Technology Co Ltd USB External Devices External Devices Users can connect the mouse and other external devices recommended by Seamaty through USB port External devices are not including in the standard configuration Caution 1 Internal parts requir...

Page 43: ...on 7 3 When the analyzer is stored for more than 3 months it should be removed from the carton case and power on for 4 hours After checking the running status place the analyzer into the carton case with direction indicated on the box 7 4 Do not stack the analyzer and do not close to the ground walls and roof 7 5 Transportation temperature 20 to 55 storage temperature 0 to 40 relative humidity 85 ...

Page 44: ...1 Tianhe Road High tech Zone Chengdu Manufacturing Address No 10 Room 1 the first floor Unit 1 Building 1 No 1 Tianhe Road High tech Zoen West Park Chengdu Tel 86 028 69760308 Fax 86 028 69760307 Post 611731 Website www seamaty com After sales Service Company Name Chengdu Seamaty Technology Co Ltd Address No 1 Tianhe Road High tech Zone Chengdu Hotline 86 028 69760306 Email sales seamaty com Instr...

Page 45: ...NOTE Please read this manual carefully before using analyzer ...