MAN.200.EN--05--SDT200-user-manual-Eng.docx
14/19
Specific technical specifications for the
SDT200 ATEX version
The SDT200 ATEX version is certified ATEX II 1 G Ex ia IIC T3/T2 Ga,
according to IEC 60079-0, 60079-11 and 60079-26.
Check room temperature for use of the SDT200 ATEX version in
potentially explosive environments. The max. permissible room
temperature is:
-15°C to 50°C (T3).
-15°C to 60°C (T2).
The SDT200 ATEX version has the following marking on the rear side:
Electrical specification for non SDT sensors used with the SDT200
ATEX version
Check electrical compatibility of non SDT sensors, for use in potentially explosive environments, before
connecting them to the SDT200 ATEX version.
Max. outer voltage V
0
20.4V
Max. outer current intensity I
0
68.7mA
Max. outer effective capacitance C
0
96nF
Max. outer effective inductance L
0
1mH