E04: Temperature error
The ambient temperature deviates from
the operating temperature range of the test
device.
Solution
Move the Analyzer and test device into
proper temperature range and perform
the test. Do not artificially heat or cool the
Analyzer.
E05: Communication error
Communication between Analyzer and
barcode or between Analyzer and printer has
failed.
Solution
If the error persists after rebooting the
Analyzer, consult the SD Biosensor.
E06: Total hemoglobin amount error
The total hemoglobin measured deviates
from the range of 7-23 g/dL.
Solution
Test again with a new test device.
If the error persists after rebooting the
Analyzer, contact the SD Biosensor.