Precautions
Caution
To reduce the risk of damage to Analyzer
·
Use the Analyzer for
in vitro diagnostics only.
·
Keep the Analyzer in a flat and dry place. Avoid direct sunlight.
·
Do not expose the analyzer to strong light. Otherwise,
serious interference to the test results may occur.
·
Do not move the Analyzer during the test.
·
Do not drop the Analyzer. Otherwise, the Analyzer may be
damaged.
·
Do not attempt to disassemble the Analyzer.
·
Do not immerse the Analyzer in water or cleaning solution.
To reduce the risk of inaccurate results
·
This Analyzer should be used only by trained operators.
·
If the error message occurs repeatedly, stop using the
Analyzer.
·
For accurate results, refer to the manual that comes with
each test device.
·
Do not use test devices which have expired.
To prevent electric shock
·
Ensure the Analyzer is completely grounded. Do not connect
the Analyzer to an ungrounded outlet.
·
When using a power board, use a grounded power board.
·
Make sure that all cables are connected to the Analyzer
correctly and completely.
·
There is a risk of electric shock if Analyzer is not grounded.
Use the power cable provided by the SD Biosensor. It
is recommended to use a power cable of 250V ~, 10A,
0.75mm2 (18 AWG).
·
Do not share power supply with other instruments and/or
devices.