2) Shake the control solution bottle and discard the first drop of solution. Gently
squeeze the bottle to form one small drop. Bring the drop to the edge of the strip, and
allow the strip to automatically draw the control solution into the yellow window. When
control solution is applied to the test strip, the meter counts down from 5 to 1 on the
display. Tightly replace the cap on control solution.
3) The control solution result appears on the display in just 5 seconds.
4) Compare the control solution result to the range printed on the test strip vial
container. If the results are not within the control range printed on the test strip
vial, then the meter and strips may not be working properly. Repeat the control solution
test. If your results are still out of range contact your supplier.
5) Remove the used test strip from the meter and discard it.