background image

 

 
 
 
 

 

 
 

 

 

 

Summary of Contents for 02LA10G

Page 1: ......

Page 2: ...ucose 32 2 Test Procedure for cholesterol lipid only 33 3 Procedure for glucose 36 4 Understanding Your Results 37 CHAPTER 5 CONTROL SOLUTION TEST 40 1 Control Solution Test 40 2 Procedure for the lip...

Page 3: ...ease read the following references U S Food and Drug Administration FDA Public Health Notification Use of Fingerstick Devices on More Than One Person Poses Risk for Transmitting Bloodborne Pathogens I...

Page 4: ...disorders such as diabetes mellitus atherosclerosis and various liver and renal diseases Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus nephrosis...

Page 5: ...HDL by an enzymatic method based on the method formulation of Allain et al 6 and Roeschlau 7 To measure TC TG and HDL in the blood the following formulations are used GLUCOSE The SD LipidoCare Profes...

Page 6: ...lyzer for glucose is designed for testing a fresh capillary whole blood sample from the fingertip palm upper arm or forearm Refer to the Safety Information There is a potential risk of infection To cl...

Page 7: ...and SD Glucose Control Solution are required but not included and must be purchased separately If you want to purchase these products please contact our customer service center at 855 486 2888 Appeara...

Page 8: ...odechip for lipid testing I Optic cover Holds the inserted test strip to prevent the movement of the test strip when you are testing J Data port Download test results to a personal computer K Printer...

Page 9: ...e Indicates measurement being performed Indicates beep sound setting Indicates the operating temperature is out of the range Indicates Lipid Profile test strip Open or close the flap Units of test res...

Page 10: ...attery warning is displayed for the first time approximately 50 measurements can still be performed In this case replace the batteries as soon as possible When replacing the new batteries you must res...

Page 11: ...partment Please note the orientation of the battery head and terminals flat end Use only alkaline manganese batteries 1 5 V AAA 5 Close the battery compartment lid 6 Turn the analyzer on to test the f...

Page 12: ...tooth PC On Off Off 13 Bluetooth Printer On Off Off 14 Bluetooth Repairing Printer On Off Off These functions are only for the 02LA20G with Bluetooth module STAGE 0 Entering the setting mode 1 Press t...

Page 13: ...t stage you can set the beep function 1 If you enter the setting mode the first step of the setting mode will appear 2 Turn the beep sound on or off by pressing either the left or right button and the...

Page 14: ...button Stage 4 Setting Date In the 4th stage you can set the date 1 After setting the date format select the actual date by pressing either the left or right button and then save it by pressing the ON...

Page 15: ...Set AM or PM Sub step 2 Set hour 1 to 12 Sub step 3 Set minute 1 to 59 b 24h Format Sub step 1 Set hour 1 to 24 Sub step 2 Set minute 1 to 59 Stage 7 Setting Unit of the Test Result In the 7th stage...

Page 16: ...s automatically right after testing Stage 9 Setting Simplex Duplex Printing In the 9th stage you can set either single sided or double sided printing options 1 After the auto printing function is set...

Page 17: ...the hypo level as 60 70 or 80 mg dL by pressing the left or right button Then save it by pressing the ON OFF button NOTE Please consult your healthcare provider for an appropriate hypo level from amon...

Page 18: ...will automatically turn off For the model 02LA20G additional setting stages remain Please refer to the next page for the additional setting steps 3 If you select ON for the first alarm the clock icon...

Page 19: ...After setting the Bluetooth with a PC the Bluetooth and printer icons will appear on the screen 2 Turn the selection on or off by pressing the left or right button and then save it by pressing the ON...

Page 20: ...test kit are the same Code Setting 1 Make sure that the analyzer is turned off Remove the old or inserted code chip from the analyzer 2 Insert a new code chip until it snaps into place Then turn the...

Page 21: ...glucose press the ON OFF button and hold it for 2 seconds Then the bt icon will appear on the left side of the screen with the setting icon which means pairing is ready 2 Open the Lipid Management Sof...

Page 22: ...one message will appear on the screen At the same time the analyzer will display the OK message on the screen If you receive both messages on the PC and the analyzer pairing was done successfully 6 Af...

Page 23: ...nly for the mode you selected just one mode at a time Indicates that it displays the test results stored in the memory of all the parameters NOTE The data will display with the corresponding parameter...

Page 24: ...ccording to the order of the testing time When entering the All mode of the memory the latest result with the largest index number will display first If you press the left button the second latest res...

Page 25: ...an review them all by pressing the left or right button The following picture describes the parameter in sequence On the screen both the and icons appear in the Lipid mode How to review the data When...

Page 26: ...sult in the analyzer 6 Displaying the average of glucose test results The analyzer can calculate 7 15 and 30 day averages of glucose test results stored in memory The HI Lo result results outside of t...

Page 27: ...alyzer 7 Deleting test data 1 Switch the analyzer on and then press the left or right button in order to enter the Memory mode at the test strip lipid or glucose standby state 2 Navigate through to id...

Page 28: ...ill not display the Printing symbol When a test is done the test result sheet is automatically printed without any action Manual Printing If you press the SET PRT button in the Memory mode the Printin...

Page 29: ...test strips 1 Getting a blood sample from fingersticks 1 Twist off the protective lancet cap 2 Position the lancet and press it firmly against the puncture site 3 Gently apply intermittent pressure to...

Page 30: ...lood glucose is not rising or falling rapidly Things to Know When using alternative sites Please read the following before testing at alternative sites palm forearm and upper arm The capillary whole b...

Page 31: ...6 Lift the disposable lancing device straight up 7 Touch and hold a drop of blood to the edge of the test strip until the yellow window is completely filled with blood The blood is drawn into the str...

Page 32: ...Do Read the test strip package insert Operate the analyzer at the acceptable test conditions For lipid testing 64 90 F 18 32 C 10 90 RH For glucose testing 50 113 F 10 45 C 10 90 RH Place the analyze...

Page 33: ...u will hear a beep 4 After the display appears open the door by lifting it Then the display below will appear 5 Prepare the blood samples collected from your fingerstick NOTE For the procedure to coll...

Page 34: ...ard it The analyzer shuts off automatically 3 seconds after the used test strip has been removed NOTE The icon will be displayed on the screen with the test reult Please press the left or right button...

Page 35: ...sample to the test strip the Close Door message will appear on the screen accompanied by warning beeps If you do not close the door within 20 seconds after the blood sample is applied the analyzer wil...

Page 36: ...ive site using a lancet and lancing device NOTE Refer to Chapter 3 Collecting Blood Samples for instructions on how to collect the blood sample from your fingerstick or alternative site 5 Touch the to...

Page 37: ...0 7 34 HDL 25 95 0 65 2 46 25 0 65 95 2 46 Glucose 20 600 1 2 33 3 20 1 2 600 33 3 For Lipid After finishing the test you can review test results by pressing the left or right button the total cholest...

Page 38: ...High 500 5 65 Very High LDL 100 2 59 Optimal 100 129 2 59 3 34 Near optimal Above optimal 130 159 3 36 4 11 Borderline high 160 189 4 14 4 89 High 190 4 91 Very High Non HDL ATP III identifies non HDL...

Page 39: ...of expected blood glucose values for you Expected blood glucose values for non diabetic adults are as follows10 Before eating 100 mg dL 5 6 mmol L One to two hours after meals 140 mg dL 7 8 mmol L Re...

Page 40: ...packaging Whenever your result does not agree with your symptoms CAUTION Use only the specified control solution SD Lipid Control Solution SD Glucose Control Solution Check the expiration date on the...

Page 41: ...solution insert If the results are not within the lipid control range printed on the insert the analyzer and the test strip may not be working properly Repeat the control test If the control solution...

Page 42: ...m 5 to 1 seconds on the screen 5 The SD Glucose Control Solution result appears on the screen in just 5 seconds 6 Compare the glucose control solution result to the range printed on the glucose test s...

Page 43: ...d retest Did you insert a lipid or glucose test strip firmly into the Analyzer Make sure a lipid or glucose test strip is inserted into the lipid or glucose test strip slot until it goes no further Di...

Page 44: ...mode of the lipid check strip press both the left and right buttons together at the strip standby mode and hold for 3 seconds Then the CHE message will appear on the screen 2 Insert the SD Lipid Check...

Page 45: ...pecial mode to enter for this glucose check strip testing If the SD Glucose Check Strip is inserted into the analyzer the analyzer starts the check strip test automatically 2 The check strip test resu...

Page 46: ...html Cleaning is the general removal of debris such as food feces urine blood saliva or other body secretions but cleaning does not kill germs effectively Disinfection is the process of killing disea...

Page 47: ...ediately if you find the following Thin silver streaks appear become cracked soft dissolved brittle or swollen clouding of LCD display You are unable to turn on off your meter or operate left right bu...

Page 48: ...your hands thoroughly with soap and water CAUTION Do not get water inside the analyzer Never immerse the analyzer or hold it under running water because it will damage the analyzer Do not spray direc...

Page 49: ...It can be discolored if the lipid check strip is exposed to sunlight Control solution 1 Do not use any SD Glucose or Lipid Control Solution that has passed the expiration date 2 For the SD Glucose Con...

Page 50: ...t is higher than the measuring range of the parameter Warning Your test result is lower than the measuring range of the parameter Warning Low Battery At this time the battery is getting low but you ca...

Page 51: ...pid test strips are expired Solution Discard the inserted lipid test strip and perform the test again using a new lipid test strip Temperature Error If the environmental temperature is above or below...

Page 52: ...during the measurement Blood Error The door is closed for testing without a blood sample Solution Re test with a blood sample Communication Error with Code Chip The analyzer cannot communicate with th...

Page 53: ...rameter TC TG 30 55 HDL 30 52 2 Test Strip for Glucose 1 The SD LipidoCare Professional System is not designed to be a substitute for pathology laboratory equipment and it should not be used for the d...

Page 54: ...mg dL 2 59 11 64 mmol L TG 45 650 mg dL 0 51 7 34 mmol L HDL 25 95 mg dL 0 65 2 46 mmol L Glucose 20 600 mg dL 1 2 33 3 mmol L Sample Lipid Fresh capillary whole blood from the fingertip Venous whole...

Page 55: ...Third Report of the National Cholesterol Education Program NCEP Expert Panel on Detection Evaluation and Treatment of High Cholesterol in Adults Adult Treatment Panel III JAMA 2001 285 2486 97 6 IDF...

Page 56: ...es Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is...

Page 57: ...20G Manufacturer SD BIOSENSOR INC C 4th 5th 16 Deogyeong daero 1556 beon gil Yeongtong gu Suwon si Gyeonggi do 443 813 Republic of Korea Importer Authorized Representative SD BIOSENSOR USA INC 2 Corpo...