Static Voltage Settings
1. Toggle the joystick until the Static Voltage channel
is displayed (Figure 13).
2. To reset the ESD counter to zero, push down the
joystick momentarily.
3. To change the voltage range and thresholds, push
down and hold the joystick until "REF" is displayed
on the top-left comer. "REF" indicates that the
settings can be modified.
4. Toggle the joystick up or down to scroll through the
voltage range options (12.5, 100,250,500, 1000
volts). Select the 12.5 volt range to enable the
ionization balance function if desired.
5. Toggle the joystick left to decrease the threshold.
Toggle the joystick right to increase the threshold.
Hold the joystick in either direction to quickly
increment the value.
6. Push down the joystick momentarily to save all
settings.
"MEM"
will appear temporarily in the top
left comer while the settings are saved. The monitor
will then re-boot after the save is completed.
NOTE: The monitor will revert back to its last saved
settings if idled for 10 seconds while in the settings
menu. The joystick must be pushed down in order
to save all modifications to the settings.
Ionization Balance and Decay Channel
Ionization
-
....!i,i
==,
-
-----
J.:
-�
..
Ionization
Balance
Balance
Polarity
, ,
J
2
Decay_____
f.
:r
•Decay Time
Measurement
.�
in Progress
* -
U
8
1
• Ionization
Ionization
-,_
Balance
Balance Level
I
_ti
.:
Threshold
Bar Graph
••• •••
•
Marker
Figure 14. Balance and Decay Mode channel
Ionization Balance Polarity:
Indicates the polarity of
the ionizer's balance (offset voltage).
Decay Measurement
in
Progress:
Appears when the
decay time is being measured.
Ionization Balance Level Bar Graph:
Indicates the
balance (offset voltage) levels.
Ionization Balance:
Displays the ionizer's balance
( offset voltage).
Decay Time:
Displays the amount of seconds to decay
a static charge. "1--.--" will display if the decay time
exceeds 16 seconds.
Ionization Balance Threshold Marker:
Marks the
balance (offset voltage) threshold setting.
NOTE: The EM Aware Monitor is compatible with
steady-state DC critical environment ionizers only. The
ionizer must have a balance ( offset voltage) of ±40V or
better.
Ionization Balance and Decay Settings
NOTE: These settings cannot be changed unless the
static voltage range is set to 12.5 volts. See "Static
Voltage Settings" on page 15.
1. Toggle the joystick until the Ionization Balance and
Decay channel is displayed (Figure 14).
2. Push down the joystick momentarily to manually
measure the ionizer's balance (offset voltage) and
decay time. The *- symbols will appear while the
measurements are being taken. The balance (upper
row) and decay time (lower row) will update at
the end of the tests. Balance and decay time are
automatically measured every three minutes. Static
voltage monitoring becomes idle when balance and
decay tests are performed. ESD event monitoring
operates normally.
3. To change the voltage range and thresholds, push
down and hold the joystick until "REF" is displayed
on the top-left comer. "REF" indicates that the
settings can be modified. The value in the top row
indicates the current balance threshold, and the
value in the bottom row indicates the decay time
threshold.
4. Toggle the joystick down to decrease the decay
time threshold. Toggle the joystick up to increase
the decay time threshold. Hold the joystick in either
direction to quickly increment the value.
d
E
[ ..............
Decay Label
3. D ◄::-- Decay Time
Threshold
(seconds)
Figure 15. Decay time threshold settings
5. Toggle the joystick left to decrease the balance
threshold. Toggle the joystick right to increase
the balance threshold. Hold the joystick in either
direction to quickly increment the value.
TB-9097
Page 15 of 19