Installation
NOVA
WAVE
should be installed on a stable lab counter top which provides easy access considering the height of the lab’s operators. The
exhaust vent of the
NOVA
WAVE
must be connected to a fume hood to prevent noxious fumes from entering the lab in the event of the
venting of a digestion vessel.
Installation Location
To install
NOVA
WAVE
select a location that:
1. Is in close proximity to lab exhaust (a proper laboratory fume hood).
2. Provides a minimum clearance of 3 in. (7.5 cm) space on all sides.
3. Provides adequate bench space for IN and OUT staging tables or
NOVA
WAVE
FA.
4. Provides a mounting surface of inflammable material capable of supporting 75 kg.
5. Is free from vibration of large equipment.
6. Provides close access to a properly grounded electrical outlet (30A, 230V, single phase)(receptacle provided).
7.
NOVA
WAVE
should be operated on a stabilized constant voltage AC power supply, and the voltage must
be /-5% of the specified level.
8. Should be lifted from bottom by 2 or 3 people.
9. Do not lift by the side panels or door openings.
NOVA
WAVE
SA
Install the IN and OUT staging tables by ensuring the tongue of the staging table connectors are installed and fixed to the bottom of the
NOVA
WAVE
frame. Level the tops by placing a straight edge inside the tunnel. Turn the adjustable feet up or down to the point that the
bottom edge of the straight edge is touching the top of the staging tables. Adjust the front feet until the side panels are level with the table top.
Connect the exhaust tubing to the exhaust port of the
NOVA
WAVE
. The system is now ready to be plugged in.