KeyPad service manual
Rev.1.1
5
freight charges for service parts purchased after the warranty
period.
4.
SCP SCIENCE
reserves the right to request that failed parts be returned
to a location specified by
SCP SCIENCE
, in which case freight charges
shall be borne by (distributor). This clause shall apply irrespective of
whether the unit is under warranty or not.
5. A
Digi
PREP
digestion system that does not operate upon receipt (‘out-of-
box failure’) may be returned to
SCP SCIENCE
with transport charges to
the account of
SCP SCIENCE
. Note that the distributor must obtain a
Return Authorization number from
SCP SCIENCE
before shipping the
instrument.
SCP SCIENCE
reserves the right to select the method of
transport.
6. (Distributor) agrees to cooperate in all respects with
SCP SCIENCE
and
will provide
SCP SCIENCE
with service failure information, serial numbers
of the digestion block and/or controller, and any other service related
information determined by
SCP SCIENCE
as being relevant to the repair
of an instrument and to the improvement of the product.
7. An RMA # is required prior to returning any item to
SCP SCIENCE
.
8.
SCP SCIENCE
shall not be responsible for any damage or losses
however caused, which may be experienced as a result of the installation
or use of this product.
1.1 ERROR MESSAGES
Some of these error messages will not appear on older equipment.
1.
Digi
PROBE
ERROR - ensure the
Digi
PROBE
or the
Digi
PROBE
Dongle
is installed correctly.
2. Over Temperature Protection - the block is over heating the set point
temperature check the SSR section 3
3. Temperature Dropping -
Digi
PROBE
is no longer in the solution and
could be measuring vapor temperature.
4. Check
Digi
PROBE
and
Digi
SET
Dongle - The set point temperature has
not been reached. But steady state has been reached.
5. Steady State Error. The system is detecting a constant temperature but
has not reached set point temperature.
6. Temperature Unsteady
– the
Digi
PROBE
appears to be sensing a
variation in temperature typical caused when the
Digi
PROBE
is or is about
to become out of the solution due to evaporation.
7. Set point Exceed Sample Boiling Point
– Verify the set point entered.
Note: Any of these error messages will be eliminated upon the pressing of any
key. Elapsed time is not interrupted in this sequence.