11.Normal Malfunction and debugging

Encountered symptoms

Possible cause

Debugging

Power on without display

1.Power supply is not connected.

Check the power supply system to see if there is voltage

2.Block have access to jack socket

Plug the equipment firmly

3.The power switch have not been turned on

Turn on the power switch on the right side of the equipment

4.The fuse is broken

Replace the fuse with new one of same specification

5.Malfunction of power box circuit occurs.

Notify the manufacturer for repair service

Equipment enter into 

the state of  temperature 

exceeding alarm

1.Equipment has not yet enter into the state of 
constant temperature

Wait a minute and observe for a while

2.Setting of the parameter is unreasonable

Refer to the operating procedure and change or re-set parameter

3.Malfunction of heating system occurs

Notify the manufacturer for repair service

Real temperature is lower 

than the setting temperature,

thus activate the lower 

temperature deviation alarm

1.Equipment has not yet enter into the state of 
constant temperature

2.Temperature deviation alarm value is too small

Wait a minute and observe for a while

Reset the alarm value

3.Abnormal condition occurs with the heater

Notify the manufacturer for repair service

Screen displays nothing 

or lack of stroke or distortion

1.Equipment is disturbed by high frequency

2.Eliminate the source of disturbance and restart the
 operation

3.Block has access to displaying circuit

Eliminate the source of disturbance and restart the operation

Refer to professionals for check or repair service

Notify the manufacturer for repair service

14

Summary of Contents for Oven-110S

Page 1: ...This Manual applies to Please read thisGuide attentively beforeusing and be careful of keepingthe Manual incase of contingentuse User s Manual Oven S Series Blast Air Oven...

Page 2: ...ire cause by electric leakage 9 5 When moving the equipment Moving the equipment connected with power supply will probably result in electric shock Danger Warnings against likely injury or death Pleas...

Page 3: ...ration clock indicating poweroff recovery parameter recalling and encryption temperature indicating and correcting RS 232 interface and built in microprinter 4 Monitoring timer independent exceeding g...

Page 4: ...eating indicator Actual Temperature indicator Time indicator Operating status indicator Setted Temperature Indicator Status keeping indicator Refrigeration indicator Operating button R H S R H S SET S...

Page 5: ...status indicator When the equipment is well functioned this indicator will be light Status keeping indicator Under program control mode when the equipment is in the state of temperature keeping this...

Page 6: ...nfirm The Ramp End Temperature RET and Hold Time HT of each segment can be set in the same way Each segment contains 3 parameter settings RT RET HT 6 Setting of parameter under the constant temperatur...

Page 7: ...ram press the R H S and hold for 3 seconds to quit the existing program first the Run indicators goes off and press R H S again to restart the whole program Case B HSD 0 1 Attention 2 SET SET SET SET...

Page 8: ...2 Setting of the clock Press the SET key and enter the password 4 The clock time can be changed when deviation between clock time and standard time occurs Press the SET key and enter the password 4 Pr...

Page 9: ...igned programme the operating indicator RUN turns on 2 When the equipment is in the state of temperature control or programme control the heating indicator lights or twinkles means the heating is conn...

Page 10: ...arameter is unreasonable Refer to the operating procedure and change or re set parameter 3 Malfunction of heating system occurs Notify the manufacturer for repair service Real temperature is lower tha...

Reviews: