Trouble Shootings 

 
If there is a need to access the (top) control panel, use an Allen wrench of 6mm (In-
cluded). Push the Allen screw in and turn at the same time. The cover will be raised a 
little 

– enough to lift it and take it out. 

 
 
 
 

 
 
 
 
 
 

Observed 
symptoms 

Possible cause 

Corrections 

 
 
 
 
Incubator does not 
work / 
No display 

1.Power supply is not 
connected 
 

Check supply system to see if there is power 
on the outlet. 

2. The power switch has not 
been switched on. 
 

Turn on the power switch on the right side of 
the incubator 

3. The (inside) fuse is broken 
 

Replace  fuse  with  new  one  of  same  specifi-
cation 

4. Malfunction of power box 
circuit occurs  

Notify distributor to repair the unit. 

 
 
Incubator temper-
ature alarm 

1.Unit has not yet reached 
the required  (constant)  
temperature 

Keep waiting and observe for a while 

2. Setting of the alarm 
parameter is wrong. 

Refer to the operating procedure and change 
or re-set alarm parameter 

3. Malfunction of heating 
system. 

Notify your distributor for repair  

 
Real temperature 
(PV) is lower than 
the set (SV) 
temperature. This 
activates  low 
temperature alarm 

1. Equipment has not yet 
entered into the state of 
constant temperature 

Keep waiting and observe for a while 

2. Temperature deviation 
alarm value is too small 

Reset the alarm value 

3. Abnormal conditions occur 
with the heater.  

Notify your distributor for repair 

4.  In  the  programme  mode, 
the difference between preset 
value  and  measured  value 
keeps changing. 

In the programme mode, turning off the alarm 
function is strongly recommended. 

Screen display 
shows nothing or 
just strokes and or 
distortions 

1.  Equipment  is  disturbed  by 
high frequency. 

Eliminate  the  source  of  disturbance  and 
restart the operation. 

2. Microprocessor failure 
 

Notify your distributor for repair 

Fluorescent  lamp 
does not work 
 

1. The lamp tube is broken. 

Replace the lamp tube 

2.  The  starter  is  loose  or 
broken. 

Check or replace

 

the starter 

 

 

 
 

 

Summary of Contents for INCU-160C

Page 1: ...User s Manual This Manual applies to Models INCU 90C INCU 160C INCU 270C INCU 430C INCU C series Economic Cooled BOD Incubator...

Page 2: ...dustries such as medical pharmaceutical and agricultural It is widely used for breeding fermentation micro organism cultivation constant temperature environmental test denaturation experiments and the...

Page 3: ...In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power supply Continuance of abnormal state will end up with fire as a result of ove...

Page 4: ...peration do not open the door too much or too long as this might affect the temperature inside 9 The incubator must be kept away from volatile flammable explosive liquids or gases 10 Please keep the c...

Page 5: ...ration and Start up 11 5 Setting of the Control Mode 12 6 Setting Parameters for Constant Temperature Control Mode 13 7 Setting Parameters for Programmable Control Mode 15 8 Settings of the Other Func...

Page 6: ...nd cool down the incubator to the required temperature The 3 dimensional air circulation motor fan speed adjustable in three steps via the con troller ensures optimum temperature distribution across t...

Page 7: ...igital Display LCD Grids Included 2 Max 11 2 Max 15 2 Max 18 2 Max 25 Grid Size mm WxD 310x356 410x456 513x556 555x656 Inner Dimensions mm WxDxH 400x410x550 500x500x650 600x600x750 700x645x950 Exterio...

Page 8: ...INCU C Series V2018 1 1 8 3 Control Panel Illustration of Control panel...

Page 9: ...g button Under fixed value mode press the button to start the unit Press the button again to stop In the Programmable mode press the button to start running the unit Press it again to enter the Hold s...

Page 10: ...this indicator will light it twinkles when the screen displays the preset time Operating status indicator When the unit operates as normal this indicator will light up Status hold indicator Indicator...

Page 11: ...wer is applied to the unit and the LCD displays the following in turn Time Function Preset Temperature Measured Temperature Clean the chamber of incubator before you put it into use and on a regular b...

Page 12: ...ired control mode Stop current operation by pressing the Operation button R H S Press the SET key and enter the password 1 Press the SET key again until the LCD display shows the control mode options...

Page 13: ...et function Press the Increase or Decrease key to set the de sired temperature 6 2Set Timer Only if you want to run the incubator on the Timer otherwise just leave it as 0 To set timer press the Set k...

Page 14: ...heating up temperature control mode model INCU 160C INCU 270C for time and temperature at an ambient of about 25 C This will help you to set the parameters in the Programme mode chapter 7 Other model...

Page 15: ...gment 1 Use the decrease and or increase buttons 7 8 Press SET to store and enter the next segment S2 setting process 7 9 Press increase or decrease button to enter the second Ramp Time S2 And so on 7...

Page 16: ...wing segments 10 in total Special attention To pause the current program press R H S key and press again to resume To abort the program any time press the R H S key for 3 seconds To continue press the...

Page 17: ...n to enter into time set menu Press Increase or Decrease key to set the right time of day Press the SET key again the display shows the year press the Increase or Decrease button to change it to the a...

Page 18: ...240 hours Set it as 0 to disable the auto defrosting function Once the defrost timer is set the unit will automatically perform the defrosting cycle accordingly To find out the suitable defrost timer...

Page 19: ...aph Press the SET key to enter the password 8 Press the Set key again to enter rec ord time function Press the Increase or Decrease key to modify the interval time for chart display Press the Set key...

Page 20: ...ed the chamber is fully loaded and forced air circulation is required this unit is installed with main ventilating fan speed adjusting option to provide further flexibility for various applications Th...

Page 21: ...R H S button once more and the unit will start again to count down from the preset operation time 9 4 While the unit is in use the current remaining operating time cannot be changed If however changed...

Page 22: ...reading is 6 C difference would be 6 8 2 C while if reading is 9 C the difference would be 9 8 1 C 11 4 Press the SET button and go with the up arrow to code 47 11 5 Press the SET again to enter the S...

Page 23: ...mperature Point Correction 100 0 11 9 If the incubator is to be used for more than one temperature setting and High Temperature Point Correction 100 0 needs to be performed as well 11 10 Change the se...

Page 24: ...ter is wrong Refer to the operating procedure and change or re set alarm parameter 3 Malfunction of heating system Notify your distributor for repair Real temperature PV is lower than the set SV tempe...

Reviews: