7” Square Hotplate

2. Proper use

The  instrument  is  designed  for  mixing  and/or  heating 

liquids in schools, laboratories or factories. 
• Observe  the  minimum  distances  between  the  devices,

between  the  device  and  the  wall  and  above  the  assembly 

(min. 100 mm) 

>100mm

>100mm

Figure 1 

This device is not suitable for using in residential areas or 

other constraints mentioned in Chapter 1. 

3. Inspection

3.1 Receiving Inspection 

Unpack the equipment carefully and check for any damages 

which  may  have  arisen  during  transport.  Please  contact 

manufacturer/supplier for technical support. 

Note: 

If  there  is  any  apparent  damage  to  the  system, 

please do not plug it into the power line. 

3.2 Listing of Items 

The package includes the following items: 

Items 

Qty

 

Main unit 

Power cable 

User Manual 

Table 1 

Summary of Contents for SCI550-H

Page 1: ... LED Digital 7 Square Hotplate Please read the User Manual carefully before use and follow all operating and safety instructions We cannot be responsible to inform at real time if the outline and specifications are subject to change for improvement ...

Page 2: ...iving Inspection 3 3 2 Listing of Items 3 4 Control 4 4 1 Control elements 4 4 2 Display 5 5 Trial run 5 6 Working with external temperature sensor 6 7 Residual heat warning 6 8 Faults 7 9 Maintenance and Cleaning 7 10 Associated standards and regulations 8 11 Specifications 9 12 Ordering information 10 ...

Page 3: ...e or software Methods and procedures adopted to resolve the problems Your contact information Warranty This instrument is warranted to be free from defects in materials and workmanship under normal use and service for a period of 24 months from the date of invoice The warranty is extended only to the original purchaser It shall not apply to any product or parts which have been damaged on account o...

Page 4: ...the media used Be aware of hazards due to unsafe container Process pathogenic materials only in closed vessels Check the instrument and accessories prior to each use Do not use damaged components Safe operation is only guaranteed with the accessories described in the Accessories chapter Accessories must be securely attached to the device and can not come off by themselves Always disconnect the plu...

Page 5: ...suitable for using in residential areas or other constraints mentioned in Chapter 1 3 Inspection 3 1 Receiving Inspection Unpack the equipment carefully and check for any damages which may have arisen during transport Please contact manufacturer supplier for technical support Note If there is any apparent damage to the system please do not plug it into the power line 3 2 Listing of Items The packa...

Page 6: ...ons Temperature control knob The heating function is switched ON or OFF by rotating knob LED Display If rotate the heating knob LED displays the temperature setting value and shift to real value in the duration of 5 seconds LED Heat When the heating function is switched ON the LED Heat flash Power Switch Switch ON or OFF the instrument Table 2 ...

Page 7: ...s Hot otherwise LED displays 0 5 Trial run Make sure the required operating voltage and power supply voltage match Ensure the socket must be properly grounded Plug in the power cable ensure the power is on and begin initializing Place vessel on the work plate Set the target temperature and start heating Stop the heating functions If these operations above are normal the device is ready to operate ...

Page 8: ...witch OFF the instrument Ensure the external temperature sensors is inserted in the media heated Switch ON the instrument and run heating function If the heating function did not work please contact manufacturer supplier for technical support 7 Residual heat warning In order to prevent the risk of burns from a hotplate digital hotplate has a residual heat warning function When the heating function...

Page 9: ...ng tenside Isopropyl alcohol Cosmetics Water containing tenside Isopropyl alcohol Foodstuffs Water containing tenside Fuels Water containing tenside Wear the proper protective gloves during cleaning of the instrument Before using other method for cleaning or decontamination the user must ascertain with the manufacturer that this method will not harm the instrument Send in the case of service the i...

Page 10: ...e equipment NOTE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in a...

Page 11: ...ment 5 Temperature display C LED Temperature display accuracy C 1 Control accuracy of heating temperature C 10 The safety temperature of the hotplate 580 Temperature sensor in medium PT1000 Control accuracy of heating temperature with temperature sensor 0 5 Residual heat warning 50 Dimensions mm 215 360 112 Weight kg 3 8 Permitted ambient temperature C 5 40 Permitted relative humidity 80 Protectio...

Page 12: ...plug 200 240V 50Hz 60Hz 51212122xxxx SCI550 H LED Digital 7 Square Hotplate glass ceramics hotplate Euro plug 200 240V 50Hz 60Hz 51212132xxxx SCI550 H LED Digital 7 Square Hotplate glass ceramics hotplate UK plug 200 240V 50Hz 60Hz Accessories 18900016 PT1000 A Temperature sensor for digital hotplate model length of 230mm 18900136 PT1000 B Glass coated temperature sensor for digital hotplate model...

Page 13: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com ...