PAGE 25
HOW TO CLEAN THE ELECTRODE
If too many residues have been found on electrode
working area, it is suggested to clean the electrode
as too many residues attached might disturb the
chemical reaction during measuring. You can do
membrane change and electrolyte refill together.
1. Rotate to unscrew and disassemble the probe.
2. Use soft/long fiber cloth to clean the “surface of
working area” where residues attached on.
NOTE: If the residues cannot be removed by cloth, please use fingernail to
scrape Iightly on the “surface of working area” to remove it.
3.
Use a knife blade to scrape lightly on the “electrode” to remove the white residue as
below image.
NOTE: Please take care on this part, do not
scrape the “surface of working area”
by knife blade to prevent damage.