SmartMeter 

 

3/4 

 

CONFIGURATION CYCLE 

To  enter  the  configuration  cycle,  press   

 

BACK 

  and 

 

SET 

.  The  meter 

displays 

typ.1

the  first  parameter  of  the  cycle.  To  access  the 

remaining parameters ofthis cycle, press 

 

SET 

.

  Keys 

 

 and 

 

 

make  possible  to  change  the  parameters  defined.  To  access  the 
measures screen, go to the end of this cycle or press keys 

 

BACK 

 and 

 

SET 

 again. 

The parameters of the configuration cycle are: 

typ.1

 

Selects the channel 1 sensor according to Table 1. Channel 1 sensor 
is chosen regardless of the channel 2 sensor selection. 

typ.2 

Selects the channel 2 sensor according to Table 1. Channel 2 sensor 
is  chosen  regardless  of  the  channel  1  sensor  selection.  Channel  2 
indication can be disabled. 

Code

 

Input

 

oFF

 

Channel 2 - disabled 

0-50 

0-50 mV 

Rtd

 

Pt100 

Tc b

 

B thermocouple 

Tc s

 

S thermocouple 

Tc r

 

R thermocouple 

Tc n

 

N thermocouple 

Tc e

 

E thermocouple 

Tc t

 

T thermocouple 

tc h

 

K thermocouple 

tc j

 

J thermocouple 

Table 1 

dP.P1

 (Decimal Point Position) 

Enables  the  display  of  decimal  points  for  channel  1  measures. 
Parameter not valid for thermocouples 

R, S 

and 

B

. For the 0-50 mV 

input, defines the decimal point position. 

dP.P2

 

(Decimal Point Position)

 

Enables  the  display  of  decimal  points  for  channel  2  measures. 
Parameter not valid for thermocouples 

R, S 

and 

B

. For the 0-50 mV 

input, defines the decimal point position. 

In.l1

 (Lowest indication limit of channel 1 for 0-50 mV) 

Determines  the  inferior  limit  of  the  user-created  meter  indication 
range,  when  the  0-50  mV  input  signal  is  selected.  Parameter  not 
displayed for the other input types. 

In.k1

 (Highest indication limit of channel 1 for 0-50 mV) 

Determines the upper limit of the user-created meter indication range, 
when the 0-50 mV input signal is selected. Parameter not displayed 
for the other input types.

 

In.l2

 (Lowest indication limit of channel 2 for 0-50 mV) 

Determines  the  inferior  limit  of  the  user-created  meter  indication 
range,  when  the  0-50  mV  input  signal  is  selected.  Parameter  not 
displayed for the other input types.

 

In.k2

 (Highest indication limit of channel 2 for 0-50 mV) 

Determines the upper limit of the user-created meter indication range, 
when the 0-50 mV input signal is selected. Parameter not displayed 
for the other input types.

 

0fs.1

 (Indication Offset, channel 1) 

Allows for an offset in the temperature value displayed in channel 1. 
The  offset  value  is  simply  added  to  the  value  measured  by  the 

sensor. The offset can be of 10 % of the selected sensor maximum 
range, upwards or downwards.

 

0fs.2

 (Indication Offset, channel 2) 

Allows for an offset in the temperature value displayed in channel 2. 
The  offset  value  is  simply  added  to  the  value  measured  by  the 
sensor. The offset can be of 10 % of the selected sensor maximum 
range, upwards or downwards. 

beep 

(Beep – Sound alarm) 

It enables or disables the sound alarm

 

Beep

 = 

no  

: disabled 

Beep

 = 

yes 

: enabled 

Note

:  The 

ALARM

  message  is  displayed  regardless  of  the  sound 

alarm status (on or off). To stop beeping while an alarm condition is 
present, press any key. 

Sh.0f

 (

shut-off

 – Auto power off) 

It makes possible to define a time interval in which the meter powers 
off automatically, provided no key is pressed within the interval. 
Adjustable between 3 and 120 minutes.

 

vnit 

(Unit – Temperature indication unit) 

It defines the temperature unit of the meter. 

ºC

 for degrees Celsius and 

ºF

 Fahrenheit.

 

Filt 

(Noise filter) 

 

It  adapts  the  meter  frequency  to  the  mains  voltage  in  order  to 
optimize  the  internal  noise  filter.  Such  adaptation  reduces  possible 
temperature oscillations, resulting from the mains frequency. Options 
are: 

50

 - for 50 Hz mains. 

60

 - for 60 Hz mains. 

Fvnc 

(Special functions) 

Defines  whether  special  functions  HOLD,  AVG,  MAX,  MIN  and 
RELAT will be applied to channels 1 and 2 or only to channel 1. 

Ch1

 -  special functions applied only to channel 1. 

Ch1

 -  special functions applied to channels 1 and 2  

4.4

 

CALIBRATION CYCLE 

To enter the calibration cycle, press 

 

BACK 

 and 

 

SET 

 for 10 seconds. The 

meter  displays 

soft

the  first  parameter  of  the  cycle.    To  access  the 

remaining  parameters  of  this  cycle,  press 

 

SET 

.

   Keys 

 

 and 

 

 

make  possible  to  change  the  parameters  defined.  To  access  the 
measures screen, go to the end of this cycle or press keys 

 

BACK 

 and 

 

SET 

 again. The parameters of the configuration cycle are: 

Soft 

(Software version)

 

About information. Informs the software review installed. 

8888 

(Serial Number)

 

About  information.  Informs  the  meter  serial  number.  The  four  larger 
numbers are the most significant and the smaller the less significant. 

L(.(1 

(

L

ow 

C

alibration Channel 1)

 

INPUT OFFSET CALIBRATION

: Adjusts the lowest limit of the input 

signal  indication  range  (offset).  In  order  to  change  the  indicated 
value, it may be required to press 

 

 or 

 

 several times. 

k(.(1 

(

High

 

C

alibration Channel 1)

 

INPUT  SPAN  CALIBRATION

:  Adjusts  the  highest  limit  of  the input 

signal indication range (span). In order to change the indicated value, 
it may be required to press 

 

 or 

 

 several times. 

(j.Lo

 

Adjusts  the  room  temperature  where  the  meter  is  held.  Important 
information for calculations and meter internal compensation. 

Reviews: