scigiene SMARTMETER Instruction Manual Download Page 2

SmartMeter 

 

2/4 

4

 

OPERATION 

The  meter  features  two  input  channels.  Channel  readouts  are 
updated two times per second. Channel 1 readouts are shown in the 
largest digits and channel 2 readouts are shown in the smallest digits, 
as Figure 07 shows. Channel 02 can be turned-off by operator. 

 

Figure 06

 — Meter front panel 

 

Figure 07

 — SmartMeter display 

4.1

 

TURNING THE METER ON/OFF 

Press 

 

 

to turn on the meter. To turn off, press for 3 seconds. 

The meter basic configuration is factory set, so it can be immediately 
used with the sensor provided. Factory configuration is: two enabled 
channels,  K-type  sensor,  one  decimal  space,  zero  offset,  sound 
alarm not enabled, auto power off after five minutes, temperature unit 
in  degrees  Celsius,  60  Hz  filter  and  channels  1  and  2  for  special 
functions. 

4.2

 

FIRST READOUTS 

Insert  the  sensor  provided  with  the  meter  in  the  terminals  of  the 
required  channel  and  take  the  sensor  leads  to  the  measurement 
location. Check the indication displayed. 
Use the keys to find special functions the meter offers. When using 
other sensors, check if factory setup is still compliant. If not, access 
the configuration cycle and change settings. 

4.3

 

SPECIAL FUNCTIONS 

HOLD FUNCTION 

It “freezes” the measure reading indication. HOLD is then displayed. It 
is  activated  by  pressing  key 

 

HOLD 

.

  To  resume  the  meter  operation, 

press key 

 

HOLD 

 again. 

MIN FUNCTION 

It  allows  the  user  to  retrieve  the 

minimum

  temperature  value 

measured  within  an  interval.  An  indicator  of  such  condition  is 
displayed  in  the  meter.  It  is  activated  by  pressing  key 

 

AVG 

MAX  MIN

 

.

  To 

resume the meter operation, press key 

 

AVG 

MAX  MIN

 

 again. 

MAX FUNCTION 

It  allows  the  user  to  retrieve  the 

maximum

  temperature  value 

measured  within  an  interval.  An  indicator  of  such  condition  is 
displayed  in  the  meter.  It  is  activated  by  pressing  key 

 

AVG 

MAX  MIN

 

.

  To 

resume the meter operation, press key 

 

AVG 

MAX  MIN

 

 again. 

 
 

AVG FUNCTION (average) 

It  displays  the 

average

  temperature  value  measured  within  an 

interval. An indicator of such condition is displayed in the meter. It is 
activated  by  pressing  key 

 

AVG 

MAX  MIN

 

.

  To  resume  the  meter  operation, 

press key 

 

AVG 

MAX  MIN

 

 again. 

Notes

:  Functions  MIN,  MAX  and  AVG  are  accessed  via  key 

 

AVG 

MAX  MIN

 

Values displayed are the ones acquired from the start of an interval of 
measurements. The interval starts after a 

RESET

 action

.

 To reset the 

meter,  press  and  hold  key 

 

AVG 

MAX  MIN

 

 

for  more  than  10  seconds.  When 

reset button is pressed, MIN, MAX and AVG values are erased and a 
new interval starts. 

RELAT FUNCTION (relative) 

When enabled, it will void the measured values and starts to indicate 
variations from such zeroed values. An indicator of such condition is 
displayed  in  the  meter.  It  is  activated  by  pressing  key 

 

RELAT 

.

  To 

resume the meter operation, press key 

 

RELAT 

 again. 

T1 – T2 FUNCTION 

When  enabled,  it  displays  the  highest  value  corresponding  to  the 
difference  between  the  value  measured  through  channel  1  and  the 
value measured through channel 2. The indication corresponding to 
channel 2 is not changed. An indicator of such condition is displayed 
in  the  meter.  It  is  activated  by  pressing  key 

 

T1 – T2 

.

  To  resume  the 

meter operation, press key 

 

T1 – T2 

 again. 

Note

: This function is only enabled when both channels are enabled 

and operating. 

ALARMS

 

The meter has an audible alarm (BEEP) that can inform the operator 
about an alarm condition. In each channel there can be two indication 
values,  when  they  are  reached  the  audible  alarm  is  triggered.  A 
visual indicator is also displayed in the meter when the indication is 
under an alarm condition. 

LOW BATTERY INDICATOR 

A low meter indicator is displayed warning that the battery should be 
replaced. 
SET UP 
To set up the meter, the user must access the Alarm and Meter Set 
up cycles, where a series of parameters must be defined.  

ALARM CYCLE 

In  the  alarm  cycle  you  find  the  setpoints  parameters  of  the  meter 
alarms. Press 

 

SET 

 

to access this cycle. The meter displays 

spk.1

the first parameter of the cycle.  To access the remaining parameters 
of this cycle, press 

 

SET 

 

again

,

 until you reach the measures screen

.

 

Keys 

 

 and 

 

 make possible to change the parameters defined. 

The parameters of this cycle are: 

spk.1

 (Highest setpoint of channel 1) 

Determines  a  temperature  value  in  channel  1  that  when  exceeded 
upwards  triggers  the  meter  alarm  (BEEP).  When  the  alarm  is 
triggered, the indication 

ALARM  

is displayed in the meter. 

sp1.1

 (Lowest setpoint of channel 1) 

Determines  a  temperature  value  in  channel  1  that  when  exceeded 
downwards  triggers  the  meter  alarm  (BEEP).  When  the  alarm  is 
triggered, the indication 

ALARM  

is displayed in the meter. 

spk.2

 (Highest setpoint of channel 2) 

Determines  a  temperature  value  in  channel  2  that  when  exceeded 
upwards  triggers  the  meter  alarm  (BEEP).  When  the  alarm  is 
triggered, the indication 

ALARM  

is displayed in the meter. 

spk1.2

 (Lowest setpoint of channel 2) 

Determines  a  temperature  value  in  channel  2  that  when  exceeded 
downwards  triggers  the  meter  alarm  (BEEP).  When  the  alarm  is 
triggered, the indication 

ALARM 

is displayed in the meter. 

 