All  keys  will  be  blocked  when  system  comes  into  counterdown  interface. 
Please restart it if you need to config parameter or select procedure. It will 
goes into  running  interface  in 3 seconds. Then press start/stop key over 2 
seconds,  you  will  see  parameter  configuration  interface.  At  last,  press 
start/stop key quickly to run it when you finished setting parameter. 

APF & RTF are enabled, while ORF is disabled at default. It will run ORF if ORF 

and APF are both enabled. Quickly press start/stop key, procedure will save the 
configuration information and exit current menu. Press start/stop over 1second, 
it will exit current menu without saving configuration information. 

 

 
4.7.2.  Automatic  recovery  after  power  failure

ORF

:

 

Enable ORF, it will get back to the procedure run 
before power failure, once when power on again.   
Disable ORF, it requires to select procedure when 
power on again. 

4.7.3.  Automatic  Running

APF

:  Enable  APF, 

automatic running last procedure when power on. 
 
Disable APF, Select procedure when power on. 

4.7 Other Functions 

Power on, then immediately press P key and Timer key at the same time, it will enters 
into Function Configuration Menu. 

 
4.7.1.

 

Automatically  keep  constant  temperature 

after finishing operation

RTF

):

 

Enable 

RTF

,  keep  constant  temperature  after 

finishing operation. 

 

Disable 

RTF

, start cooling after finishing operation. 

APF              DISABLE 

APF              ENABLE 

ORF            DISABLE 

ORF              ENABLE 

RTF              ENABLE 

RTF              DISABLE 

Summary of Contents for Mycotox CUSFY-1

Page 1: ...1 Operation Manual for Mycotoxin LFD Incubator CUSFY 1...

Page 2: ...stallation 5 2 Technical Parameter 6 3 Safety Warning 6 4 Operation Procedure 7 4 1 Set up a new Procedure 7 4 2 Set S N of Procedure 7 4 3 Set Temperature 7 4 4 Set Time 7 4 5 Set Time Unit 8 4 6 Pro...

Page 3: ...oods and contact us immediately if you found any parts mismatched missing or damaged 1 Introduction Incubator is new product controlled by microcomputer It is widely used for sample preservation and r...

Page 4: ...4 1 2 Primary Structure Metal Module Display Operation Panel Power Switch Power Input Socket...

Page 5: ...r input socket then connect the other end of power adapter with power grid between 100V and 240V Check below chart 1 3 3 Power switch on and then it will reach to setting temperature Basically it take...

Page 6: ...y 0 1 Heating up Time from 40 to 100 20min Environment Temperature 5 35 Overall Dimension W D H 110mm 148mm 120mm Net Weight 0 5 kg 3 Safety Warning It is used indoor It might cause personal injury if...

Page 7: ...em comes to operation interface Press Start Stop key over 2 minutes then is goes into Setting up Procedure interface You can DIY 9 procedures P1 to P9 Each procedure includes 3 sections move A B C 28...

Page 8: ...o A Stop key 4 6 6 It will send warning tone at one minute at a time and also display FINISHED after running complete procedure Meanwhile it will remain at current temperature Please turn off the powe...

Page 9: ...will exit current menu without saving configuration information 4 7 2 Automatic recovery after power failure ORF Enable ORF it will get back to the procedure run before power failure once when power o...

Page 10: ...ntact us Module did not heat up Heater malfunction Contact us Key malfunction Broken membrane switch Contact us Regularly clean longhole on module with clean soft cloth that takes a little absolute al...

Reviews: