User Manual SM6025
23
The following control parameters can be set on
<BIN NO. DISP>
.
Compare function ON/OFF (COMP)
There are 2 zones: BIN NO. DISP, COMP. Their detailed information will be introduced as below.
The following test conditions are displayed in the measurement result/condition zone. These zones
cannot be set on this page but can be set on
<MEAS SETUP>
,
<MEAS DISP>
or
<CORRECTION>
.
Test function (FUNCA)
Test frequency (FREQ)
Test level (LEVEL)
Test range (RANG)
Test speed (SPEED)
DC BIAS
ON/OFF set state of OPEN, SHORT, LOAD (CORR)
3.2.1 Comparator function
SM6025 has an inserted compare function which can divide DUT to up to 10 bins (from BIN1 to BIN9
and BIN OUT). Users can set 9 pairs of primary parameter limit and one pair of secondary bin limit. If the
primary parameter of DUT is within the range of the bin limit but the secondary parameter is outside of
the bin limit, the DUT will be sorted into the auxiliary bin. When SM6025 has installed an HANDLER
interface, the compare result will be output into the automatic test system and further realizing
auto-sorting test. These limits can only be set on the
<LIMIT TABLE SETUP>
page. Users can set the
compare function to ON or OFF in the
COMP
zone.
Operation steps for compare function
:
1) Move the cursor to
COMP
, the following soft keys will be displayed.
ON
OFF
2) Select one of above soft keys to set the compare function as ON or OFF.
3.3
<BIN COUNT DISP>
Press [DISP] and then select the soft key of BIN COUNT to enter into the
<BIN COUNT>
page which
shows the count of each bin.