Cal 9500 Configuration Data 

 

Level 3                                                                  Level P 
SP1.d – rly                                                            Prog – 1 
SP2.d – rly                                                            Run - OFF 
Burn – uP.SC 
rEV.d – 1d.2r 
rEV.L – 1i2n 
SPAn – 0.0 
Zero –  3 (Note: Only Change this parameter if calibrating unit, value    

        

may vary.) 
ChEk – off 
rEAd  - Var 
tECh – CtA 
Ver – 953 
rSET – none 
 
Level 2 
SP1.P – 0 
hAnd – off 
PL.1 – 100 
PL.2 – 100 
SP2.A – Dv.hi 
SP2.b – none 
di SP – 0.1 
hi .SC – 100 
Lo. SC –  -100 
in Pt – tc t 
unit – C 
 
Level 1 
tune – off 
band – 0.3 
int.t – off 
dEr.t – off 
dac – 0.5 
cyc.t – on.of 
ofSt – 0.0 
SP.Lk – off 
SPrr – 0 
SPrn – off 
SoAk – -- 
Set.2 – 6 (alarm deviation) 
bnd.2 – 0.1 
Cyc.2 – on.of 

 

Summary of Contents for 45-01A

Page 1: ...ase button Provide at least 3 of space around the freezer for ventilation Avoid placing the freezer in locations exposed to direct sunlight heat registers or any other heat source Alarm On Off switch should remain in the Off position until the freezer reaches desired set point temperature When the temperature has been reached turn the alarm switch to the On position Cabinet should not be loaded wi...

Page 2: ...es not have a Chart Recorder Open up to the compartment behind and Next gently separate and slide the right grill panel off you will see two more screws DO NOT remove the left side 2 3 Remove these screws 4 Slide this part completely off to the right 5 After the grill is removed you will see the black condenser Being careful not to bend the steel fins vacuum vertically up and down ...

Page 3: ... 517 263 5492 Website www scientemp com OPERATING AND INSTALLATION MANUAL MODEL 45 01A This cabinet has passed the QUALITY CONTROL INSPECTION and meets the high standards at Scientemp Corp This inspection includes the complete refrigeration system cabinet construction and finish ...

Page 4: ...his product only in the manner described in the product literature and in this manual Before using it verify that this product is suitable for its intended use Do not modify system components The cabinet must be properly grounded in conformity with national and local electrical codes Never connect the unit to overloaded power sources Disconnect the unit from all power sources before cleaning troub...

Page 5: ...anel with chart recorder Remove screw located on right side middle of panel Instrument panel will hinge open Time delay has a dial and is located on lower right corner of electric box NOTE It is recommended to keep the freezer operating continually rather than turning the freezer off and on 4 4 Chart Recorder Optional If your freezer has a chart recorder installed On initial start up you must repl...

Page 6: ...rmer areas of your body If possible immerse your hands in water that is slightly above normal body temperature about 100 F or 37 8 C or which feels warm to someone else until your normal color returns Do not use direct heat Seek medical attention as quickly as you can 7 REFRIGERATION COMPARTMENT The refrigeration system of a Scientemp freezer is hermetically sealed It requires no maintenance Howev...

Page 7: ...play Setpoint value or program selection Green LED Setpoint 1 output indicator Upper Red LED Setpoint 2 output indicator Lower Red LED Setpoint 3 output indicator Green Display Process variable or Function Option ADJUSTMENTS To enter or exit program mode Press together for 3 seconds To scroll through functions Press or To change levels or options Press together or together To view setpoint units P...

Page 8: ...arameter if calibrating unit value may vary ChEk off rEAd Var tECh CtA Ver 953 rSET none Level 2 SP1 P 0 hAnd off PL 1 100 PL 2 100 SP2 A Dv hi SP2 b none di SP 0 1 hi SC 100 Lo SC 100 in Pt tc t unit C Level 1 tune off band 0 3 int t off dEr t off dac 0 5 cyc t on of ofSt 0 0 SP Lk off SPrr 0 SPrn off SoAk Set 2 6 alarm deviation bnd 2 0 1 Cyc 2 on of ...

Page 9: ... or by fire or flood CONSEQUENTIAL DAMAGES IN NO EVENT SHALL SCIENTEMP CORP BE LIABLE FOR SPECIAL INDIRECT OR CONSEQUENTIAL DAMAGES INCLUDING WITHOUT LIMITATION PRODUCT LOSS OR PRODUCT SPOILAGE CLAIMS NOR FOR ANY DELAY IN THE PERFORMANCE OF THIS WARRANTY DUE TO CAUSES BEYOND ITS CONTROL GENERAL The standard warranty and any service contract related to the STANDARD WARRANTY shall apply only to the ...

Page 10: ...CK WHITE CAL 9500 CONTROLLER 6 5 4 3 2 1 N L INPUT TC COMPRESSOR RECORDER RED POWER RELAY RED WHITE BLACK WHITE WHITE RED BLACK BLUE RED BLACK BLACK L1 BLACK BLACK BLACK WHITE ORANGE ORANGE BLACK WHITE RED RED WHITE WHITE L2 BLACK WHITE RTD RED RED RED RED BLACK RED RED RED 8 7 45 01 WIRING DIAGRAM CAL 9500 WHITE FAN ...