30
*WAI
This is the IEEE488.2 common command. From either interface, it causes the command processor to wait
until any overlapping command is finished before continuing to process commands.
CLR
This command halts RPT?, SND?, or COL? commands in progress. It also terminates statistics gathering
started by the COL command. This command is issued internally whenever a carriage return that is not
preceded by a command is typed into the RS232 interface. When the IEEE488 interface receives a linefeed
terminated message with no commands, it too generates the CLR command.
*RCL_<dec num>
This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the
instrument configuration to that configuration stored in the corresponding save area. It also places the IEEE
interface into the OCIS and OQIS states.
COUNT_<dec num>
Sets the number of pulses making up each average in average energy mode.
COUNT?
Returns the number of pulses making up each average in average energy mode.
CALIB_<dec num>
Sets the calibration constant (Volts/Joule) for a pyro detector, (Volt/Watt) for an Ultra Series detector or the
time constant for a calorimeter. Note: This command is context sensitive, the detector type must be set by the
SENS command before this command is executed.
CALIB?
Returns the constant for a pyro detector, or the time constant for a calorimeter. Note: This command is
context sensitive, the returned value depends on the detector type set by the SENS command.
ATTEN_<dec num>
Sets an attenuation factor for the current detector.
ATTEN?
Returns the attenuation factor for the current detector.
ZERO
This command zeros the power baseline for a calorimeter.
PSWD_<dec num>
Enables password protection commands if the password is entered correctly.
PSPEED HF|BL
Sets the electronic response speed according to the repetition rate capability of the pyroelectric detector type -
a "black" coated absorbing crystal, or uncoated "high frequency" absorbing crystal.
PSPEED?
Returns the pyro speed, possible responses are BL and HF.
CSPEED_<dec num>