M2/M2 G4 Handpiece Adaptor (LCS) Operator Manual
Coupling
MULTIflex - p/n 01-113930
or
INTRAmatic - p/n 01-113931
with adaptor
(not included)
Luer Lock Fittings
Handpiece
(not included)
Connecting handpieces to the LCS adaptor
Insert the handpiece coupling into the handpiece.
Push the handpiece into one
of the ports.
1
2
2
3
Multiflex
Intramatic
Push tab
to release
Program settings
To ensure maximum wash performance results of the
handpieces and the other instruments, the Technician
needs to increase the wash time to the maximum setting
of 15 minutes for the cycle.
Note:
The overall cycle time will increase accordingly by
about 6 minutes.
1
Please note: As recommended by leading manufacturer/suppliers of handpieces, the hollow parts of instruments should be dried immediately after cleaning and disinfection using
medical compressed air before additional processes or sterilization. This is to be followed by recommended regular maintenance procedures and application of maintenance
agents/oils. Please follow all applicable health & safety regulations for your region.
Using the touchscreen menu
The HYDR
IM
will
automatically detect
the presence of the
LCS adaptor. Select
the desired program
and follow the prompts.
Upon completion, open
the door and remove
the handpieces from the
adaptor by pushing on the
tab at the side of the port.
Example: Regular cycle
Please note: As recommended
by leading manufacturer/suppliers
of handpieces, the hollow parts
of instruments should be dried
immediately after cleaning and
disinfection using medical compressed
air before additional processes or
sterilization. This is to be followed by
recommended regular maintenance
procedures and application of
maintenance agents/oils. Please
follow all applicable health & safety
regulations for your region.
1
2