Maintenance and troubleshooting

LCS system Messages          What it Means 

                                        What to Do – User

CLEANING THE LCS WATER FILTER

REPLACING THE BACTERIOLOGICAL AIR FILTER

Return the filter to the 
housing and close the 
cap tightly. 

Remove the filter.

Rinse the filter and, if required,  
clean with a soft brush or using  
compressed air. Check and clean  
the O-ring. Remove any debris  
inside the housing.

Unscrew the cap at  
the front of the  
LCS adaptor cylinder.

Replace the bacteriological filter  
every 500 cycles or if dirty.
To change the filter:

1.  

Power the unit OFF,  
disconnect tube A and remove  
the filter from the bracket.

2.  

 Disconnect tube B from the  
back of the filter.

3.  

Before installing the replacement filter  
(01-102119S) check that the arrow mark 
on the filter matches the direction of  
the arrow on the bracket. (Arrow should  
be pointing to the right).

4.  

Insert filter into bracket and connect  
tube B. Re-connect tube A.

5.  

After replacing the filter, go to the “Reset 
LCS Counter” screen in the Setup menu 
and reset the counter to 0.

Replace Water Filter 

Replace Air Filter 
 

 

CF 12 Water  
Filter Clogged 

 

CF 18 Water Filter  
Pressure Sensor Stuck

CF 26 Air Valve error 

CF 30 LCS adaptor  
disconnected

Water filter in the LCS adaptor  
is dirty or clogged.

The bacteriological air filter in the  
kickplate is dirty.
 

Problem with Lumens Cleaning System. 

 

 
Problem with Lumens Cleaning System.  

Problem with air suppy pressure in air line.  
Incoming air pressure must be between 1 bar 
(15 psi) - 10 bar (145 psi).

Problem with Lumens Cleaning System.

•  Follow instructions above:  

“Cleaning the water filter.”

•  Follow instructions above:  

“Replacing the air filter.”

•  Check LCS adaptor is correctly positioned.
•  You selected LCS at program start but  

LCS is not installed.

•  Replace Water Filter, see  

“Cleaning the water filter.”

•  If cleaning doesn’t resolve the problem,  

replace the water filter (01-113742S) 

•  If message persists, call for service. 
 
•  Call for service. 

•  Call for service. 

•  Check that LCS adaptor is correctly positioned.

Tube A

Tube B (behind)

1

2

4

3

English

Summary of Contents for Hydrim C61wd G4

Page 1: ...LCS Operator s Manual Mode d emploi Bedienerhandbuch Manuale d uso Manual del operador 95 113906 EU ML R4 HYDR IM C61wd G4 Lumens Cleaning System LCS Installation Guide and Operator s Manual Copyrigh...

Page 2: ...plug on the inside back left of the chamber 1 2 3 English Multiflex Intramatic Push tab to release Setting up the LCS unit Remove the screw and pull out the white plug on the back wall of the unit s c...

Page 3: ...e A 5 After replacing the filter go to the Reset LCS Counter screen in the Setup menu and reset the counter to 0 Replace Water Filter Replace Air Filter CF 12 Water Filter Clogged CF 18 Water Filter P...

Page 4: ...hambre 1 2 3 French Multiflex Intramatic Appuyer sur l attache afin de lib rer Mise en place de l adaptateur LCS Retirer la vis ainsi que l l ment enfichable blanc de la paroi ar ri re de la chambre d...

Page 5: ...tion et r initialiser le compteur sur z ro Remplacer filtre eau Remplacer filtre air Filtre eau CF 12 bouch Capteur de pression filtre eau CF 18 bloqu Erreur vanne d air CF 26 CF30 LCS Adaptateur d co...

Page 6: ...Waschkammer 1 2 3 German Multiflex Intramatic Dr cken Sie zum L sen die Lasche Montage des LCS Adapters Entfernen Sie die Schraube und ziehen Sie den wei en Stopfen auf der R ckwand der Kammer des Ge...

Page 7: ...ter ausgetauscht haben gehen Sie im Setup Men in den Bildschirm LCS Z hler zur cksetzen und setzen den Z hler auf null zur ck Wasserfilter austauschen Luftfilter austauschen CF 12 Wasserfilter verschm...

Page 8: ...2 3 Spanish Multiflex Intramatic Presione la pesta a para soltar Gradilla para instrumentos verticales opcional Ajuste del adaptador de SLL Quite el tornillo y extraiga el tap n blanco situado en la p...

Page 9: ...raci n y vuelva a colocar en cero el contador Reemplazar filtro de agua Reemplazar filtro de aire CF 12 Filtro de agua obstruido CF 18 Sensor de presi n del filtro de agua atascado CF 26 Error en v lv...

Page 10: ...bianco 1 2 3 Italian Multiflex Intramatic Premere la linguetta per lo sgancio Montaggio dell adattatore SPL Togliere la vite ed estrarre il tappo bianco presente sulla parete posteriore della camera d...

Page 11: ...el menu Setup e resettare il contatore a 0 Sostituire filtro acqua Sostituire filtro aria CF 12 Filtro acqua intasato CF 18 Sensore pressione filtro acqua bloccato CF 26 Errore valvola aria CF 30 Adat...

Page 12: ...LCS LCS HYDRIM C61wd G4 LCS LCS LCS HYDRIM LCS 1 2 3 LCS HYDRIM C61wd G4 HYDRIM LCS 1 2 3 Chinese Multiflex Intramatic LCS LCS HYDRIM LCS LCS LCS LCS 1 2 4 3...

Page 13: ...LCS LCS O LCS 500 1 A 2 B 3 01 102119S 4 B A 5 LCS 0 CF 12 CF 18 CF 26 CF 30 LCS LCS 1 bar 15 psi 10 bar 145 psi LCS LCS LCS LCS LCS 01 113742S LCS A B 1 2 4 3 Chinese...

Page 14: ...SciCan GmbH Wangener Strasse 78 88299 Leutkirch GERMANY Tel 49 0 7561 98343 0 Fax 49 0 7561 98343 699 SciCan Medtech AG Alpenstrasse 16 6300 Zug SWITZERLAND Tel 41 0 41 727 7027 Fax 41 0 41 727 7029 0...

Reviews: