English
15
The TENS eco 2 switches OFF automatically after two minutes if no key is pressed. In
this case none of the adjusted parameters are saved.
Memory Query Mode
First set the date and time (see chapter setting date and time).
The memory query mode is used to read out saved therapy data
from patients. By pressing and holding the right
key and
key for about 3 seconds you enter the memory query mode. If
no values are saved, the display shows “null” for about 2 seconds
and the device returns to standby mode automatically. If you
have saved therapy data, the display shows the parameters of
the last therapy session as shown below:
• Used program = P01
• Therapy duration = 0 min 44 s
• Frequency = 100 Hz
• Day of therapy = 5 (fifth day)
• Pulse width = 200 μs
• Therapy session no. = 3
You can save up to 90 therapy sessions respectively the last 3 of each day.
»
With the left
key you can return to the values of the previous day (until you reach the
first recorded day).
»
The left
key lets you proceed to the next day (until you reach the current or last day).
»
The right
key brings you to the parameters of the previous therapy session of the
respective day (until you reach the first session of the respective day).
»
The right
key shows you the parameters of the next therapy session (until you reach
the last session of the respective day).
Resetting the memory
Enter the memory query mode (see above). Pressing and holding the E key for about 5
seconds will delete all saved data. The display shows “Clr” for about 2 seconds and the
device can be used for stimulation again. The TENS eco 2 switches OFF automatically
after two minutes if no key is pressed.