schwa-medico Pointoselect Digital DT Instruction Manual Download Page 8

 

Pointoselect Digital DT   |   Instruction Manual 

 

Frequency range (partly divided through 128) 

Nogier

 

Bahr

 

Frequencies of meridian 

by Reininger/Bahr

 

2.3 Hz 

4.6 Hz 

6.44 Hz  lung 

4.6 Hz 

9.4 Hz 

4.32 Hz  large intestine 

9.1 Hz 

18.7 Hz 

3.68 Hz  stomach 

18.3 Hz 

37.5 Hz 

5.48 Hz  spleen/pancreas 

36.5 Hz 

74.9 Hz 

3.88 Hz  heart 

73.0 Hz 

149.9 Hz 

6.18 Hz  small intestine 

146.0 Hz 

2.3 Hz 

5.21 Hz  bladder 

 

 

 

 

4.77 Hz  kidney 

 

 

 

 

4.14 Hz  circulation 

 

 

 

 

10 

5.72 Hz  triple warmer 

 

 

 

 

11 

4.55 Hz  gall bladder 

 

 

 

 

12 

3.45 Hz  liver 

 

Adjusting the pulse width 

Press the left 

+ / -

 keys of the device to set the pulse width. The pulse width can be 

adjusted in steps of 10 µs from 60 to 120 µs. 
 
 
 

 

Start the stimulation 

Stimulation can be carried out in every operating mode. 
Press the 

key of the hand probe to stimulate between the inner and outer 

sensors. The parameters and the intensity used are the ones set up previously in 
the S mode.  
The letter „S“ for stimulation, the intensity in V and the frequency in Hz are 

displayed on the screen. The red LED of the hand probe is lighting up.  
The intensity can be modified during stimulation pressing the left 

+ / -

 keys on the device or the 

+ / -

 keys 

of the hand probe. The frequency can be modified during stimulation using the right 

+ / -

 keys on the 

device. 
 

Adjusting the volume of the acoustic signal 

Press and hold the 

M

 and the left 

-

 key and switch the device ON with the 

 key 

to access this mode.  
Adjust the volume level from 0 to 8 by pressing the left 

+ / -

 keys. The acoustic 

signal is switched off with “0” and is at his maximum with “8”. 

The symbol  

  indicates that the signal is deactivated. 

 

Switching OFF the device 

Press the 

 key to switch off the device. If the battery voltage falls below the critical value or the device 

detects no valid measurements over a two-minute period and no stimulation is started, the device 
switches off automatically. 
 

Summary of Contents for Pointoselect Digital DT

Page 1: ...Pointoselect Digital DT Digital Point Finder for Auricular and Body Acupuncture With Integrated PuTENS Stimulation Item no 200506 Instruction Manual Item no 100858 ...

Page 2: ...surement 6 Automatic Point Detection 6 Manual Point Detection 7 Gold silver Point Detection 7 Stimulation 7 Adjusting the volume of the acoustic signal 8 Switching OFF the device 8 Changing the battery 9 Short Instruction to find an ear point 9 7 General Information 9 Adjustments Modifications and Repairs 9 Warranty 9 Guarantee 9 Maintenance and Cleaning 10 Classification 10 Information on Battery...

Page 3: ...r or on children Always store the Pointoselect Digital DT in its case to protect it from damage and dust Warning The Pointoselect digital DT should not be used close to other electronic devices e g mobile phones in order to avoid interference If this is not possible the device must be observed during operation to ensure that the respective functions are operating correctly The application of the P...

Page 4: ...matic and manual location functions and integrated electrical stimulator Power supply 9 V battery Dimensions approx 115 x 60 x 30 mm Weight approx 123 g without accessories Point finder Current consumption approx 17 mA without signal tone approx 30 mA with signal tone Stimulator Current consumption approx 28 mA under 20 kΩ load at 128 Hz and 120 μs Output voltage 200 V under 20 kΩ load Frequency r...

Page 5: ...the Pointoselect digital DT Take particular care of the connecting pins of the hand probe while plugging the hand probe in and out Note that the hand probe can only be connected to its socket in one direction Follow the arrow on the hand probe and on the socket of the device The patient needs to hold the hand electrode and the acupuncturist needs to touch the patient Starting the device Start the ...

Page 6: ...the hand probe are required for measurements Body Exclusively the inner sensor is used for measurements in the body mode Gold and silver point detection is not active and therefore not displayed Measurement The lower the sensitivity is few bars on the display and deep acoustic signal the more relevant the point is for treatment Many bars on the display and high acoustic signal symbolize high sensi...

Page 7: ...cted the mode Go only gold points will be detected and displayed If you have selected the mode Si only silver points will be detected and displayed Please Note Silver and Gold points can only be detected if you have selected an mode Stimulation Entering the stimulation mode The stimulation settings can be modified in the S mode Press the M key to access the stimulation mode S Following parameters ...

Page 8: ...d outer sensors The parameters and the intensity used are the ones set up previously in the S mode The letter S for stimulation the intensity in V and the frequency in Hz are displayed on the screen The red LED of the hand probe is lighting up The intensity can be modified during stimulation pressing the left keys on the device or the keys of the hand probe The frequency can be modified during sti...

Page 9: ... probe at a 90 degree angle to where you want to find a point on the ear Put the probe near a point and apply a little pressure If you can reduce the sensitivity display and sound at least to half of the original value then you are exactly on the point If you choose a manual mode select with M key you have to reduce the sensitivity slowly with the minus key on the probe until you have reached at l...

Page 10: ...n points you re contributing to protecting our environment and the earth s natural resources Batteries are made from important resources and chemicals including lead cadmium zinc lithium and mercury If batteries are disposed of as normal municipal household waste they ll be taken to a landfill site and those resources will be lost and will contribute to the pollution of the environment Recycling i...

Page 11: ... with the accessories listed in the chapter Delivery Content of this manual The manufacturer cannot guarantee the security of the device if not used with original accessories 8 Delivery Content Item no Description Quantity 200506 Pointoselect digital DT 1 200507 Hand probe 1 250200 Hand electrode 1 602000 Battery 1 106242 5 24 cable 1 106752 Belt clip 1 100858 Instruction manual 1 ...

Page 12: ...struction Manual Manual version 09 2014 Distributor schwa medico GmbH Wetzlarerstr 41 43 35630 Ehringshausen Deutschland www schwa medico com export schwa medico de Pierenkemper GmbH Am Geiersberg 6 35630 Ehringshausen Deutschland ...

Reviews: