Technical check
The manufacturer recommends a technical check on the AS SUPER 4 digital every
24 months. This includes:
1. Confirming that user instructions have been included in the accompanying
documentation.
2. Checking the equipment for completeness.
3. Visual check:
- for mechanical damage
- for damage to all cables and sockets
4. Functional Safety
- Checking the output signals with a load of 1 kΩ real current and voltage
- Checking the frequency
- Checking the pulse width
These technical checks shall only be performed by qualified persons. The results must
be documented with the date and name of the person who carries out the check.
Re-adjustments, modifications and repairs
The manufacturer is only responsible for the safety and performance of the AS
SUPER 4 digital when readjustments, alterations and repairs are carried out by
authorized persons and when the unit is used in accordance with the operating
instructions.
Short instruction manual
»
key to switch ON / OFF the device
»
P
key to select the program group (1 to 9)
»
E
key to select the subprogram
»
Press
P
key for 4 seconds to enter the selecting mode for a second program
»
Press the
▲
key to increase the intensity
»
Press the
▼
key to decrease the intensity
»
Press the
E
key to pause the stimulation
»
Press the
P
key to stop the simulation
»
Press the
E
key for 4 seconds to enter the set up mode
»
for free programs (Group 1 and 3)
»
Press the key for 4 seconds to delete the changed parameters
»