1 Intended Use
The AmpliMove medical device has been designed for the generation of transcutaneous
stimulation of muscles and nerves with mid-frequency current on humans and may
not be used for any other purposes.
2 For safe use
Safety Instructions
Please read the operating instructions carefully before using the AmpliMove medical.
Health disorders may indicate serious disease. Please consult a physician and/or a
physiotherapist and consider the contra-indication and side-effects of the AmpliMove
medical.
To ensure safe use of the AmpliMove medical:
»
Only use the AmpliMove medical with original accessories.
»
Keep the AmpliMove medical away from water or other liquids.
»
Do not use the AmpliMove medical if it is not working properly or if it has been
damaged in any way.
»
The USB connector serves exclusively for software updates. Do not connect any
other device (e.g. printer) than a USB flash drive with a software update to the
USB connector of the AmpliMove medical device. Do not try to charge an external
appliance with the USB connector. Do not treat a patient with AmpliMove medical
when using its USB connector.
»
Do not simultaneously connect a patient to AmpliMove medical and to.
»
Do not operate the AmpliMove medical close to a high frequency surgical device,
another medical unit, a short wave or microwave unit or a mobile telephone system
as it can cause fluctuations in the output values of the electric current stimulation
unit.
»
Avoid electrodes placement over the anterior chest wall. The electrodes should be
positioned in such ways that the current flow will not cross through the heart.
»
The sizes of the electrodes are to be not less than 2 cm². If the applies current density
goes below 2 mA/cm², special attention on the patient is required.
»
Only connect one patient to the AmpliMove medical.
»
Do not use the AmpliMove medical during the simultaneous operation of other
machines, when driving or when asleep.
»
Do not hold anything in the hand if using the stimulation sleeves STIMEX-TEX on
the upper extremities. Remain seated if using the stimulation sleeves STIMEX-TEX
on the lower extremities.
»
Do not drop the AmpliMove medical, do not use it inappropriately or expose it to
extreme temperatures or high levels of humidity (not less than 10°C or more than 40