Guarantee
The manufacturer issues a guarantee of 12 months from the date of purchase.
The guarantee does not apply in the following cases:
»
Damage due to improper handling
»
Defects the customer is aware of on the date of purchase
»
Damage caused by the customer
»
For wearing parts and consumable supplies like, for instance, battery, cables and
electrodes.
Maintenance and Cleaning
No special cleaning or care agents are required for the AmpliMove medical. Clean the
stimulator, the cables and the mains adapter with a soft, lint-free cloth. Please ensure
that no moisture permeates the stimulator. If moisture does permeate the stimulator,
a technical check must be carried out before re-use.
Technical Check
We recommend a technical check on the AmpliMove medical every 24 months.
This includes:
1. Confirming that user instructions have been included in the accompanying
documentation.
2. Checking the equipment for completeness.
3. Visual check:
» The type plate, the adhesive label and the inscriptions are complete
and readable
» For mechanical damage
» For damage to all cables, mains adapter, plugs and sockets
4. Functional Safety
» Checking the output signals with a load resistance of 500 Ω real current
and voltage
» Checking the frequency
» Checking the modulation depth
5. Measurement after EN 62353:2008
» Equipment leakage current
» Applied part leakage current
These technical checks may only be performed by suitably qualified persons. The
results must be documented with the date and name of the person carrying out the
check.