Operation
16
02.00 | FT-AXIA 80 EtherCat | Assembly and operating manual | en | 1009369
5 Operation
The following section provides information required when using
software to operate the EtherCAT Sensor. Communicating with the
EtherCAT sensor requires knowledge of EtherCAT standards and
operation.
5.1 Sensor Enviroment
NOTICE
Damage to the outer jacketing of the transducer cable could en-
able moisture or water to enter an otherwise sealed transducer.
•
Ensure the cable jacketing is in good condition to prevent
transducer damage.
NOTE
Transducers may react to exceptionally strong and changing elec-
tromagnetic fields, such as those produced by magnetic reson-
ance imaging (MRI) machines.
To ensure proper operation, the IP64 rating of the transducer must
match or exceed the transducer’s environment.
5.2 Sample Rate
The power-on default sample rate is 975 Hz. The “Sample Rate”
field
Object 0x7010: Control Codes
[
}
26] controls the current
sample rate. The following table lists the rounded and exact
sample rates.
Rounded Sample Rate
0.5 kHz
1 kHz
2 kHz
4 kHz
Exact Sample Rate
487 Hz
975 Hz
1990 Hz 3900 Hz