page 20
Accessories U-3003
Accessories
Cable length
Manufacturer, description
Power cord
max. 2.0 m
Various manufacturers
WARNING:
The use of accessories not listed here on the U-3003-E / U-3003-E can lead to reduced
interference immunity or to increased electromagnetic radiation.
WARNING:
PORTABLE HF-communication devices (radios) (including their ACCESSORIES such as antenna
cables and external antennas) should not be used within 30 cm (12 inches) of the parts and wires of the U-3003,
as specified by the MANUFACTURER. Failure to comply may lead to a reduction in the performance of the device.
WARNING:
The characteristics of this device, determined by EMISSIONS, permit its use in industrial environments
and hospitals (CISPR 11, Class A). When used in a residential environment (which typically requires Class B,
according to CISPR 11), this device may not provide adequate protection from radio services. If necessary, the user
must take corrective measures such as transposition or reorientation of the device.
WARNING:
Use of the U-3003 next to other devices or with other devices in stacked form should be avoided, as this
could result in inaccurate operation. If use in the manner described above is nevertheless necessary, the U-3003 and
the other devices should be monitored to ensure that they are working properly.
Guideline and manufacturer’s declaration - electromagnetic emissions
The U-3003-E / S is intended for operation in the environment as specified below. The customer or
user of the U-3003-E / S should ensure that it is operated in this type of environment.
Interference emission
Conformity
Electromagnetic environment -
definitions
RF emissions according
to CISPR11
Group 2
The U-3003 generates internal RF
energy and conducts this to the
transducer head to produce aerosol.
The RF emission is very low.
Electronic equipment in the immediate
vicinity could, however, be subject to
interference.
RF emissions according
CISPR11
Class A
The U-3003 is used in professional health
care facilities. These can be, for example :
- Hospital rooms in hospitals or clinics
- Operation rooms or intensive care units
in clinics
- Hospital rooms in nursing homes
- Medical practices
Emissions of
harmonics according
to IEC 61000-3-2
Class A
Emissions of voltage
fluctuations / flicker
according to
IEC 61000-3-3
In conformity