pH
7.1
General information
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample.
You have the following options to measure the temperature:
Automatic temperature measurement with
– Temperature sensor (NTC30 or Pt1000) integrated in electrode
(if available).
– the integrated temperature sensor of another electrode that is
connected and immersed in the sample at the same time.
– an external temperature sensor (NTC30 or Pt1000).
Manual determination and input of the temperature.
The display of the temperature indicates the active temperature mea-
suring mode:
Temperature
sensor
Resolution of
the temp. dis-
play
Mode
yes
0.1 °C
Automatic with temperature
sensor
-
1 °C
Manual