10.3
Determining the cell constant (calibration in control
standard)
Why determine the cell
constant?
Due to aging, the cell constant slightly changes. As a result, an inexact
measured value is displayed. Calibration determines and stores the
current value of the cell constant.
Thus, you should calibrate at regular intervals. The calibration data of
non-ID sensors is stored in the meter. For ID sensors, the calibration
data is stored in the sensor.
You can determine the cell constant of the conductivity measuring cell
in the ranges 0.450 ... 0.500 cm
-1
, 0.585 ... 0.715 cm
-1
or 0.800 ... 1.200
cm
-1
by calibrating in the 0.01 mol/l KCl control standard or set it man-
ually in the range 0.250 ... 2.500 cm
-1
or 0.090 ... 0.110 cm
-1
.
Besides, the fixed cell constant 0.010 cm
-1
can be selected.
In the delivery condition, the calibrated cell constant of the meters is set
to 0.475 cm
-1
(conductivity measuring cell LF413T(-ID).
CalClock
The expiry of the CalClock reminds you to clean the measuring cell. It
is still possible to measure.
The CalClock is set to 180 days in the factory. You can change the
interval (see page 142).
Stability control
During calibration, the stability control is automatically activated.
Display of calibration
data and download to
interface
You can display the data of the last calibration (see page 168 and
page 160). Subsequently, you can transmit the displayed calibration
data to the interface, e. g. to a PC, with the
<PRINT>
key.
Note
The calibration record is automatically transmitted to the interface
after calibrating.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......