Conductivity
ProLab 4000
142
ba75646d03
12/2012
10.2.3
Measurement settings
The following measurement settings are possible for conductivity mea-
suring cells:
Menu item
Setting
Description
Measuring
– Cell constant
–
Calibrated cell constant
Measuring cells whose cell constant
is determined by calibration in KCL
control standard (calibration ranges
see page 157).
The currently valid cell constant is
displayed in the status line.
–
Adjustable cell constant...
Any measuring cells with freely
adjustable cell constants in the range
0.250 to 25.000 cm
-1
.
–
Adjustable LF 313T...
Measuring cell LF313T(-ID), nominal
cell constant 0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 to 0.110 cm
-1
.
–
LF 213T
Measuring cell with a nominal cell
constant of 0.010 cm
-1
.
The cell constant is permanently set.
–
Temperature coefficient
–
TC nLF
Nonlinear temperature compensa-
tion
–
TC Lin...
0.000 ... 3.000 %/K
Coefficient of the linear temperature
compensation.
–
TC off
No temperature compensation.
–
TC nLin1...
...
TC nLin4...
Measurement with entered or deter-
mined temperature coefficient.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......