Operating manual 810353-BALDE-150220
Version 20.02.2015
Page 11 of 36
The plug connection must meet IP54 at least.
The feedback contacts (RM) are designed for 24 V / 1A. They must on-
ly be connected to safety extra low voltage systems.
To improve the immunity in industrial networks the use of a
4% 2A
line
choke is recommended in the 400V supply lines. E.g. Block transformer
LR3
40-4/2
(14mH)
When using a 2A line choke the supply line must be fused with maximum
2A
.
When using several generators these can also be supplied via a common
4% choke.
The rated current of the choke must account for 2A per gen-
erator.
Maximum 16A for 8 generators.
4.4.2.
Ground connection
The ground connection POAG is used for closing the high-voltage current circuit be-
tween the high-voltage generator and deposition electrode:
CAUTION
The ground connection (Fig. 1/3) must be contacted with a POAG
socket (see Figure) and connected with min. 4 mm
2
as short as
possible with the operating ground of the ground electrode.
This connection should not be laid parallel to power supplies or
fieldbus lines.
Description: Angle socket for equipotential bonding POAG
crimp connection
Item no.:
260066
CAUTION
The ground connection does not replace the circuit breaker
(PE on power plug)
As short a possible
Ground connection (POAG)
Loading electrode Deposition electrode
HV generator