9
When you shift the tension meter from the right to the left hand, you can rotate the
readings on the DISPLAY by 180°.
Measuring with the display unit in the left hand:
If you would like to use the left hand for measuring, you should reverse the readings on
the DISPLAY to make them easier to read.
Requirement:
Tension meter switched off as described in Chapter 3.3.2.
To reverse the display:
- Press and hold the DAMP and POWER
keys until the DISPLAY shows the rea
dings the other way around.
Measuring with the display unit in the
right hand:
Requirement:
Tension meter switched off as described in Chapter 3.3.2.
To restore the default orientation:
- Press and hold the DAMP and POWER
keys until the DISPLAY shows the rea
dings in the default orientation.
3.3.4 Reversing the Display
The selected display orientation remains stored in the KXE memory even after
the instrument is switched off.
fig. 3.3.4a
fig. 3.3.4b
3.3.5 Zero Adjustment of the Measuring Position
Before starting measurement you need to carry out zero adjustment, as
described below, each time the tension meter is switched on. This procedure is
necessary to compensate for the weight of the measuring roller in the measuring
position. The zero adjustment for the new measuring path only remains effective
until the instrument is switched off.
Zero adjustment must be repeated whenever the material path is changed or the
tension meter does not display “0.”
The process material must not yet be inserted!
Requirements:
- Tension meter switched on as described in Chapter 3.3.2.
To carry out zero adjustment:
- Holding the SENSOR at the HANDLE, place it in the desired measuring position.
Be careful to hold the instrument absolutely steady.
- Press the POWER key.
The DISPLAY momentarily shows
0 0 0 0
and then switches to
.
The KXE is now adjusted for the new material path and is ready to measure.