Operation Manual Flow Transmitter DW-D
w corresponds to button
and
– to button
. The device returns back automatically
after some time to the modus set under
DWD MODE
(point 4.2) if no button is pressed. The
actual point of the menu is always shown in the first line, the chosen value in the second line.
4.1 Function
BATT CHECK
The BATT CHECK function is used for giving the user information that the device was not in
use for a certain time due to power interruption and therefore the amounts of flow could not
get accumulated during that time by the totalizer. Therefore the device should always be on
(ON) when using the totalizer function. Please find further information under point 4.6.
If the totalizer is not used, we would advise to switch-off this function (OFF), as otherwise the
device will only work after restart in case of power loss.
The menu can be operated with P and the functions with
or
. To take over the requested
adjustment please push P again.
4.2 Adjusting the relais
Press button P until
RELAIS 1
appears. Line 2 shows the actual switch point.
Pressing
or
the wished switch point can be adjusted. For setting the switch point P has to
be actuated * is appearing. After some time the sigh is disappearing. Adjusting the switch
point of
RELAIS 1
is now completed.
Actuate P button once more causes changing to
RELAIS
2.
Adjustment is like described under
RELAIS 1.
When the adjusted value is reached, a lamp glows for control. The green one (A) is assigned
to
RELAIS 1
the red one to
RELAIS 2 (B)
4.3 Adjusting of output occupancy
DW-D comes with current, voltage and frequency output. The different output types only can
be used alternatively. It is only one pair of connector existing for all functions. Connection is to
block KL1 (see point 3.2 , 3.3 and 3.4 respectively drawing of the board). At menu
OUTPUT
analogue, voltage or frequency output can be activated
.
With the buttons
and
you can set
the wished function. Button P takes over the adjustment.
At the option frequency output the max. frequency can be adjusted between 125-32000 Hz.
Therefore utton P must be pressed after entering the frequency function. Then the max.
frequency can be adjusted with the buttons
and
. Now the adjusted max. frequency
corresponds to the upper range value.
4.4
Adjusting of standard indication mode
It is possible to choose between 2 different indication modes.
At the point in the menu
DWD MODE
you can select the modes. In the second line the valid
mode is shown.
FLOW
means indication of momentous flow.
TOTAL
means that the value of
totalizer is shown. You can choose the required indication by using button
or
. Button P
takes over the adjustment.
Only the selected standard indication is permanent visible. All other points of the menu turn
back to the standard indication after some time automatically.