6
Maintenance
6.1
Maintenance Intervals
Page 40
A
rt.
no.
: 0-4
8-02
40 Rev. c
DefiSign Life AED
6.1.1
Exemption from the technical safety inspection
Exemption from the 3-year technical safety inspection is possible if the
DefiSign Life
AED
®
is exclusively used within the optimal conditions as stated below:
Optimal factors
Fulfilled
Not fulfilled
• Environmental conditions before use:
– Temperature b15...25 °C
– No daily temperature variation over 10°C
– Protection against direct sunlight
– Humidity 30 to 65 % (no condensation)
– Protection against dust
• Operation sites
– no mobile operation sites (e.g train, car, bus, airplane, …)
– not placed on walls with risk of vibrations (e.g. near doors, windows, ...)
Exemption from the technical safety inspection of the
DefiSign Life AED
if all factors are fulfilled
Yes
No
Location:
Date:
Carried out by:
6.1.2
Service/Shelf life
Device
The device has defined Service Life of 10 years if maintenance intervals have been
observed according to section
6.1 Maintenance Intervals and the directive IEC/EN
62353.
Battery
Main battery (approx.6 years), see expiring date on the battery and internal battery
cell (approx. 6 years)
Electrodes
Electrode packaging (2 years), see expiring date on the electrodes pouch.