2.4 Function
Status LED
Service
LEDs
Immediately after a battery has been inserted, the
DefiSign Life AED
®
performs a
test of the device and battery. If this test is completed successfully, the green status
LED is blink-ing and all service status LEDs are off, showing that the device has not
detected an error.
If a problem is detected during this test:
• an acoustic alarm is issued,
• the status LED stops blinking
• additional information are given by the service LEDs
Fig. 2.1
LED indicator
• If an alarm is in progress (visual and/or acoustic), the battery autonomy is reduced.
• In addition, the device performs a daily or weekly self-test (this setting must only
be configured by service personnel authorised by Schiller)
• An alarm (visual and/or acoustic) can only be reset by removing and reinserting the