Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
54
http://www.schambeck-sfd.com
■
NOTE
There are two adjustment screws for the mirror. Keep in mind that you might have to turn
the other screw. Anyway you should turn the screws not more that ½ or ¾ turns.
Adjust the screws until the optical balance reaches a value of 0 ± 10 %%.
Now the adjustment for the used solvent is finished.