Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
48
http://www.schambeck-sfd.com
To return to the former display press the
MENU
button. To exit the service mode select the
menu NormMode by pressing the
MENU
button and press
ENTER
.
16.2 E
RROR
M
ESSAGES
In case of some malfunction the RI2000 refractive index detector may display an error
message. The most error will appear directly after switching on the instrument.
16.2.1
M
ALFUNCTION OF THE
T
EMPERATURE
S
ENSOR
There are two possible error messages indicating problems with the temperature sensor. A
more detailed description of the temperature control of the RI2000 can be found in
chapter
The message “noTS” (no temperature sensor) will appear for that reason that the
temperature sensor is not connected to the circuit board or not working.
The message “!” is shown if the temperature of the optical bench is outside the range
between 9 °C and 64 °C. Beside the displayed message an acoustic signal will occur.