SO-801 Hand-Held Slit Lamp
User Manual
INSTRUCTIONS AND SPECIFICATIONS
Please read the following information carefully. Scan Optics is
responsible for the safety, reliability and performance of the equipment
only if it used in accordance with these instructions.
USE OF THE EQUIPMENT
1.
Turn the instrument on with the lower (black) knurled ring.
2.
Select the appropriate aperture. The slit lamp has three apertures
which can be selected by turning the upper (silver) knurled ring.
These apertures are:
•
Disk
: 3 mm diameter. This provides a high intensity beam for
maximum reflected scatter.
•
Slit
: 0.3 mm x 12 mm. This is used for examination of the
corneal stroma.
•
Rectangle
: 0.75 mm x 2.1 mm. This is used for examination of
the corneal surfaces. It also provides separated images of the
corneal and iris/lens surfaces for the detection of anterior
chamber flare.