 User Manual 

 

LAMP REPLACEMENT

 

The lamp is a quartz halogen type with a rated life of 22 hours. 

Replacement lamps can be obtained from Scan Optics. To replace the 

lamp, unscrew the optical head from the handle then gently unscrew the 

knurled silver cap on  

the end of the handle and remove the existing lamp. Insert the new lamp 

by holding it by the base and dropping it into the hole in the bottom of the 

silver cap. By holding the handle and cap upside down, the risk of 

dropping the lamp may be minimised. Screw the cap back onto the 

handle. Ensure that the cap is not fastened excessively tightly, so that 

damage to the lamp contacts may be avoided. 

 

Do not touch the envelope of the bulb with bare fingers, as it may reduce 

the life of the bulb. If it is accidentally touched, clean with a degreasing 

agent such as acetone or ethanol. 

 

Version 6.0 

 

Page 11 of 12 

 

Summary of Contents for SO-801

Page 1: ...SO 801 HAND HELD SLIT LAMP SO 801 ...

Page 2: ... User Manual Scan Optics University of Adelaide Research Precinct 32 Stirling Street Thebarton SA 5031 Adelaide Australia Tel 61 8 8234 9120 Fax 61 8 8234 9417 E mail admin scanoptics com au Home Page http www scanoptics com au ...

Page 3: ...ts listed are present In the case of any discrepancy advise Scan Optics immediately STANDARD PARTS Slit lamp head Battery handle Spare lamp Battery charger Spare battery User Manual this manual OPTIONAL SPARE PARTS SO 805 Spare Battery Pack SO 803 Lamp Version 6 0 Page 3 of 12 ...

Page 4: ...er Manual CONTENTS Use of Instrument 5 Battery Handle 8 Recharging the Battery 8 Battery Charging Safety 8 Battery Safety 9 Battery Capacity 9 Replacing the Battery 10 Lamp Replacement 11 Specifications 12 Version 6 0 Page 4 of 12 ...

Page 5: ...d ring 2 Select the appropriate aperture The slit lamp has three apertures which can be selected by turning the upper silver knurled ring These apertures are Disk 3 mm diameter This provides a high intensity beam for maximum reflected scatter Slit 0 3 mm x 12 mm This is used for examination of the corneal stroma Rectangle 0 75 mm x 2 1 mm This is used for examination of the corneal surfaces It als...

Page 6: ...eyepiece Look through the eyepiece Rotate the eyepiece cell to move it forwards or backwards in its thread until the slit is clear Take care not to touch the surface of the eyepiece lens The eyepiece lens is already correctly adjusted for an emmetrope when shipped from Scan Optics For a myope the eyepiece cell should be further forward that is the eyepiece lens should be closer to the slit image F...

Page 7: ...rument is brought to the examiner s eye first 6 Select the appropriate filter for the projected light by rotating the thumb wheel The options are Red free filter This filter appears green Blue filter This filter is suitable for viewing areas which have been treated with Fluorescein drops No filter Please ensure that the filter is correctly engaged in position a definitive click will be heard other...

Page 8: ...ost One or two batteries may be recharged at the same time The battery recharge time is 3 hours regardless of whether one or two batteries are being recharged In order to prolong battery life do not recharge the battery for more than three hours BATTERY CHARGING SAFETY Scan Optics SO 805 batteries and SO 804 battery chargers should only be used in conjunction with each other Do not charge the batt...

Page 9: ...be disposed of by incineration If battery liquid gets in your eyes flush them with clean water for a minimum of 10 minutes and seek immediate medical attention The battery liquid is a 25 35 solution of potassium hydroxide BATTERY CAPACITY The capacity of modern nickel cadmium batteries is not often affected by the battery discharge memory and so it is not necessary to fully discharge the battery e...

Page 10: ... the battery should first be fully discharged and recharged two or three times as described above as the cause may be reversible However if full capacity is not restored the battery must be changed Replacement batteries can be obtained from Scan Optics To replace the battery pack unscrew the cap at the bottom of the Battery Handle The new battery pack should be inserted by pushing firmly into the ...

Page 11: ...the new lamp by holding it by the base and dropping it into the hole in the bottom of the silver cap By holding the handle and cap upside down the risk of dropping the lamp may be minimised Screw the cap back onto the handle Ensure that the cap is not fastened excessively tightly so that damage to the lamp contacts may be avoided Do not touch the envelope of the bulb with bare fingers as it may re...

Page 12: ...fication 6 25 X Apertures Disk 3 mm diameter Slit 0 3 x 12 mm Rectangle 0 75 x 2 1 mm Filters Cobalt blue red free Slit angle 35 left or right LAMP Type Halogen Rated life 22 hours Voltage 3 6 V Power 5 83 W Colour temperature 3300 K MSCP 7 8 BATTERY PACK Type NiCad Capacity 45 minutes 1 2 Ah Version 6 0 Page 12 of 12 ...