SO-5800 Ophthalmic Microscope
User Manual
Page 13 of 31
Issue number 2.0
Reference SO-5800-UM-2
X-Y UNIT
The X-Y unit is positioned below the rotating joint at the end of the pantograph
arm. This unit allows the user to move the microscope head laterally, by means
of the joystick on the foot control. To move the head to the right, simply move
the joystick to the right, to move the head forward, move the joystick forward,
and so on. Movement combinations may also be achieved by moving the
joystick diagonally. Speed of X-Y movement is controlled using the knob on
the main power supply panel.
The X-Y unit also acts as an interface for distributing power to the microscope
head and video camera as well as the X-Y unit itself. These connections are
labelled on the rear panel of the X-Y unit. All voltages in this region are
<15Vdc
RETURN TO CENTRE
SENSITIVITY
ADJUSTMENT
HEAD POWER OUT
CONNECTION
VIDEO CAMERA
POWER OUT
CONNECTION
MICROSCOPE
POWER IN
CONNECTION
X-Y UNIT REAR PANEL