SAFETY MECHANISM
The SO-5800 is fitted with a safety stop mechanism that enables the user to pre-
set a particular height below which the pantograph arm will not move. Where
the knob position is set closest to the horizontal arm end, there is no impediment
to movement. When the knob is locked closest to the microscope head end,
there will be almost no movement allowed in the pantograph arm below its rest
(UP) position. Locations in between will allow a limited rage of movement.
Note that the knob MUST be locked in position for the safety mechanism to
take effect.
UNLOCK
KNOB
SLIDE INTO
POSITION
LOCK
KNOB
PUSH KNOB IN