background image

The monitor has adopted your data and 
is now ready for taking measurements. 
Pressing the ON/OFF button starts the 
automatic measurement. The cuff inflates 
to approx. 190 mmHg pressure.  

 

The automatic deflation starts the 
measurement. After about 3 seconds, a 
heart icon starts blinking in the LC 
display. If a Korotkoff’s sound is detected 
already at that time, it means a systolic 
pressure of more than 150 mmHg. Then 
the monitor re-inflates (to about 220 
mmHg). 
At the end of the measurement, the result 
is displayed in the LCD and the cuff 
deflates.  

 

Aborting a measurement 

You can stop the measurement at any 
time by pressing the ON/OFF button. The 
BP monitor deflates the cuff instantly.     

 

The memory function 

The KP-6166 saves up to 50 
measurements in every memory area. 
Together with each measurement 
(systolic, diastolic and heart rate), the 
date and time are also saved. The 
measurement with the longest residence 
time in the memory is assigned number 
1. All other measurements increment by 
one until number 50. Then the memory is 
full. The next measurement occupies 
memory location 50. The measurement 
formerly at location 50 changes to 
location 49 (etc., 48

Æ

47 .... 2

Æ

1). The 

oldest measurement formerly at location 
1 is lost. 

 

You can delete the measurements in the 
memory by removing the batteries for at 
least 2 minutes from the BP monitor. 

 

 

- 6 - 

 

WHO-Interpretation of blood pressure 
readings 

 29.12.2007   

 

Meaning Systolic 

mmHg 

Diastolic / 
mmHg 

optimal 

< 120 

< 80 

normal 

< 130 

< 85 

high-normal 

130 - 139 

85 - 89 

mildy 
hypertonic 

140 - 159 

90 - 99 

hypertonic, 
medium 
severity 

160 - 179 

100 - 109 

severly 
hypertonic 

> 180 

> 110 

   

The World Health Organization (WHO) 
recommends medical treatment, when 
the blood pressure value exceeds 
140/90 mmHg. The doctor should also 
be consulted when the blood pressure is 
too low, i.e. when the systolic value 
drops below 105 mmHg and/or the 
diastolic value below 60 mmHg. 

 

Cleaning and maintenance  

 Do not expose the BP monitor to 

extreme temperature, moisture, dust or 
direct sunlight. 

 The cuff contains a sensitive airtight 

bulb. Treat it with care and avoid any 
stress by torsion or kinking. 

 Clean the BP monitor with a soft, dry 

cloth. Do not use fuel, thinner or other 
solvent. Remove specks on the cuff 
carefully with a wet cloth and soap 
water.  

Do not wash the cuff. 

 Do not drop the BP monitor or treat it 

with force.  

 Do not try to open the BP monitor. 

This will make the factory calibration 
invalid. 

 
 

 

 

- 7 - 

 

Troubleshooting and remedies 

If a fault is detected during a 
measurement, the BM monitor displays 
„Err“ and a symbol (code for service 
personnel).  
From many years of experience we know 
that most faults are the consequence of 
wrong use. The most frequent causes of 
faults are excessive movement during the 
measurement, wrong application of the 
cuff, use of wrong batteries or use of a 
rechargeable battery. 
Wait at least 5 minutes before repeating 
the measurement. 

 

 

 

Message 

 

 

Error description 

 

Correction 

 

 

 
The pressure 
measured was lower 
than 20mmHg. 

 

 
Please measure again. 
 

 

 

 
The pressure 
measured was 
higher than 
300mmHg.  

 

 
Please measure again. 
 

 

 

 
Pumping error. 
 

 
Please check cuff and try 
again. 

 

 

 

The pressure can 
not be measured 
due to signal noise. 

 
Please measure again. 
 
 

 

 

 

The pumping 
pressure is higher 
than 300mmHg. 

 

 
Please measure again. 

 

 

or no  

display 

 
Low batteries. 

Please check and 
replace the batteries if 
necessary. 

 

Referenced norms 

For registration, a defined number of blood 
pressure monitors was subjected to 
stringent clinical testing. These monitors 
used the same program for blood pressure 
measurement as your monitor uses. Thus, 
this program is also regarded as clinically 
tested. 

 

 

- 8 - 

 

The manufacture of the blood pressure 
monitor complies with the European norms 
for blood pressure monitors: EN 60601-1, 
EN 1060-1 to EN 1060-3. 

 

Electromagnetic compatibility:

  

The blood pressure monitor complies with 
the requirements of European norm EN 
60601-1-2.  
The requirements of EU directive 
93/42/EEC concerning medical devices, 
class IIa are complied with. 

 

Warranty 

We grant a warranty of two years the for 
blood pressure monitor KP-6166. Within 
this period, defects of the BP monitor will 
be repaired free of charge. You can also 
have your BP monitor repaired by our 
customer service after the warranty period. 
We will provide you with a free cost 
estimate for the repair.   
The warranty does not cover damage due 
to improper handling, accident, natural 
wear and tear (cuff), non-compliance with 
the instructions for use or unauthorized 
modifications of the monitor. 
Excluded from the warranty are:  
the packaging, batteries and the 
instructions for use. 

 

Periodic re-calibration 

We recommend periodic checks of the 
pressure indication every 2 years. 

 

 

Disposal note  

The disposal of batteries is regulated. 
Spent batteries are collected at collection 
points provided by battery dealers and the 
communities. 

 
 

- 9 - 

Dispose of the blood pressure monitor 
as prescribed. i.e., at a collection point 
for electrical scrap. 
 

Specifications                              

Measuring method:    Oscillometric  
Measuring range:        
                         Pressure: 20 to 280 mmHg 
                         Heart rate: 40 to 195 / min 
Measuring accuracy:   
                         stat. pressure: +/- 3 mmHg 
                         Heart rate: +/- 5 % 
Measuring value memory:  2 memory 
areas, each with 50 locations for systolic /  
       diastolic / heart rate, date and time  
Display:                    LC display 
Pressure pick-up:    Semi-conductor   
                                pressure sensor 
Automatic switch-off:  after approx. 3  
                                 minutes  of inactivity  
Voltage supply:  2 x LR03 (alkali batteries), 
Working temperature range: 
                                    +10°C  -  +40°C 
Storage temperature range:   
                                   - 10°C  -  +50°C 
Weight (without batteries):     approx. 140 g  
Cuff length:  for wrist circumference from  
                    135 mm to 195 mm        

 

Specifications may change as technical 

improvements take place

 

 

 