SBS SBS-H2 Installation, Operating,  & Maintenance Instructions Download Page 2

SBS-H2 Users Manual 

Rev 1.4 

January 6

th

 2013 

Table of Contents: 

 

Description   

 

 

        Page 

 

Section 

Overview 

 

 

 

 

 

 

    1.0

 

Benefits 

 

 

 

 

 

    2.0

 

How It Works  

 

 

 

 

    3.0

 

Specifications 

 

 

 

 

    4.0

 

Sensor 

 

 

 

 

 

    5.0

 

Mounting Location   

 

 

 

    6.0 

Installation 

 

 

 

 

 

    7.0

 

Operation 

 

 

 

 

10 

 

    8.0

 

Electrical Testing 

 

 

 

10 

 

    9.0

 

Calibration 

 

 

 

 

10

 

 

  

10.0 

LED Trouble Shooting Chart

                    

11                   11.0

 

 

Part Numbers & Options: 

 

 

Part No. 

 

Description 

SBS-H2

 

 

Hydrogen Gas Detector (Includes: Main Unit, (1) sensor & (1) 25ft cable) 

H2-JB 

 

 

Junction Box, 4 11/16” X 4 11/16” Metal, with Knockouts 

H2-SENSOR

   

Additional/Replacement Hydrogen Gas Sensor with 25ft Cable 

H2-100FT-CABLE

 

Custom 100 Ft Cable and mA Converter (no sensor) 

H2-CALKIT

 

 

Field Calibration Kit for SBS-H2 

E190399

 

 

110vac 3 Prong AC Cord, 10 Ft, 18AWG 

Warnings: 

  This detector is added protection, not a substitute, for prudent safety measures where 

hydrogen gas is present. For large or highly sensitive areas, you may have to use two or 
more detectors with 1 or 2 sensors each for increased detection. 

  The hydrogen sensor does not provide protection from fires or hydrogen explosions.  

The relay contacts are intended to be connected to a safety system, enabling audible 
alarms, system shutdown, ventilation, or other measures to ensure safe handling and 
use of hydrogen gas. 

  Ensure that your installation complies completely with all relevant local, State, Federal, 

and OSHA safety and health regulations 

  If sensor responds, there is a risk of combustion or explosion.  To avoid injury leave area 

immediately. 

  The hydrogen sensor is calibrated for operation in air. Tampering with the sensor or 

operation in other gas environments can lead to inaccurate readings and possible 
permanent damage if operating in a reducing atmosphere or 100% hydrogen. The 
hydrogen sensor is silicone resistant but uncured silicone compounds or extended 
exposure to silicone off gassing, or high concentrations of refrigerant gasses related to a 
leak, can give inaccurate readings in the sensor. 

 

Pre-March 2016 Version

Summary of Contents for SBS-H2

Page 1: ...rooms and other areas where hydrogen gas may be present INSTALLATION OPERATING MAINTENANCE INSTRUCTIONS Protects Life Property and Profits 1 800 554 2243 www sbsbattery com TEST sbsbattery com Compli...

Page 2: ...have to use two or more detectors with 1 or 2 sensors each for increased detection The hydrogen sensor does not provide protection from fires or hydrogen explosions The relay contacts are intended to...

Page 3: ...ive limit LEL of 4 1 2 0 Benefits In addition to protecting your employees and your property the detector may also reduce the following costs Electricity Heating Air Conditioning Instead of continuous...

Page 4: ...t supply to the unit Relays 1 Warning Relay 1 Normally Open 1 Normally Closed contact 2 Alarm Relay 1 Normally Open and 1 Normally Closed contact 5 0 Sensor The sensor consists of a ceramic sensing el...

Page 5: ...gen gas concentration in the air immediately surrounding the sensor Hydrogen gas may accumulate in several areas of the battery compartment or room therefore individual detectors should be installed i...

Page 6: ...2 alarm relay that is energized at 2 concentration of hydrogen gas The relays can be used to switch on a remote exhaust fan and or to turn on an alarm For 120 VAC power use 18 gage stranded wire minim...

Page 7: ...le SBS Part H2 100FT CABLE the Sensor jumper must be moved on the sensor input from the off condition to the on condition so it covers both pins 1 Removing the front snap on cover will reveal the inne...

Page 8: ...er Select between single or multiple sensor set up default jumper settings single sensor Jumper on P9 Only Place a jumper on P8 to disable SENSOR 1 Place a jumper on P9 to disable SENSOR 2 B Signal Ou...

Page 9: ...ion you would like to use the relay to be related to Use Relay C for the warning condition 1 H2 and relay D for the alarm condition 2 H2 3 Restore the front covering of the alarm box and re secure the...

Page 10: ...alarm will stop and the 2 relay will de energize When the concentration falls below 1 the 1 relay will de energize and the Yellow LED will turn off 9 0 Electrical Testing after the input power is on...

Page 11: ...11 SBS H2 Users Manual Rev 1 4 January 6th 2013 Pre March 2016 Version...

Page 12: ...ther application intended to support or sustain life or for any application in which the failure of the SBS product could create a situation where personal injury or death may occur Should buyer purch...

Reviews: